Articles | Volume 2, issue 1
https://doi.org/10.5194/mr-2-117-2021
https://doi.org/10.5194/mr-2-117-2021
Research article
 | 
12 Apr 2021
Research article |  | 12 Apr 2021

Overhauser dynamic nuclear polarization (ODNP)-enhanced two-dimensional proton NMR spectroscopy at low magnetic fields

Timothy J. Keller and Thorsten Maly
Abstract

The majority of low-field Overhauser dynamic nuclear polarization (ODNP) experiments reported so far have been 1D NMR experiments to study molecular dynamics and in particular hydration dynamics. In this work, we demonstrate the application of ODNP-enhanced 2D J-resolved (JRES) spectroscopy to improve spectral resolution beyond the limit imposed by the line broadening introduced by the paramagnetic polarizing agent. Using this approach, we are able to separate the overlapping multiplets of ethyl crotonate into a second dimension and clearly identify each chemical site individually. Crucial to these experiments is interleaved spectral referencing, a method introduced to compensate for temperature-induced field drifts over the course of the NMR acquisition. This method does not require additional hardware such as a field-frequency lock, which is especially challenging when designing compact systems.

Dates
1 Introduction

In recent years, dynamic nuclear polarization (DNP) has become a robust tool to boost the signal intensity of nuclear magnetic resonance (NMR) experiments. The method has found widespread application in the areas of structural biology, imaging, and materials science. Besides increasing NMR sensitivity, DNP can be further used to obtain dynamic or spatial information about the system under study – information that is often complementary to other methods (Ardenkjaer-Larsen, 2019, 2016; Corzilius, 2018; Griffin et al., 2019; Jaudzems et al., 2019; Kaminker, 2019; Liao et al., 2018; Maly et al., 2008; Plainchont et al., 2018; Rankin et al., 2019; Rosay et al., 2016).

Currently, the majority of DNP–NMR experiments are DNP-enhanced solid-state NMR (ssNMR) experiments often under magic angle spinning (MAS) conditions. These experiments are typically performed at cryogenic temperatures (< 100 K), and commercial equipment is readily available (Rosay et al., 2016). In contrast to solution-state DNP experiments, DNP-enhanced ssNMR experiments require less demanding instrumentation since the sample can be directly irradiated by the microwave radiation without the use of a resonator. Due to the large cooling power of the cold nitrogen gas used to spin the samples for MAS experiments, the sample temperature can be kept low. Only minimal microwave-induced sample heating is observed because of the low dielectric losses of a frozen aqueous sample (tan (δ)< 0.01) (Nanni et al., 2011; Rosay et al., 2016).

For solution-state DNP experiments, the large dielectric losses of a liquid sample (e.g., water; see Fig. 4 in Neumann, 1985) present a large challenge to high-field, solution-state DNP experiments. Particularly, in the case of aqueous samples, direct irradiating of the sample with microwave power results in excessive and often destructive sample heating. To avoid sample heating, a microwave resonator is required to separate the microwave-induced electric fields, responsible for sample heating, from the microwave-induced magnetic fields required to drive the DNP process (Poole, 1967).

Many microwave resonator types and structures are known from electron paramagnetic resonance (EPR) experiments. However, since the overall dimensions of an EPR resonator scale with the wavelength of the required microwave radiation, resonators become physically very small at conventional NMR frequencies (Maly et al., 2008; Poole, 1967). For example, DNP at a proton Larmor frequency 400 MHz requires microwave radiation at 263 GHz, corresponding to a wavelength of < 1.4 mm. The resulting resonator geometries are difficult to fabricate and sample, and resonator handling is very difficult.

In addition, an radio frequency (RF) coil is required to detect the NMR signal which ideally should be located inside the microwave resonator, to assure a large NMR filling factor. If the RF coil is located outside the resonator, the gain due to DNP is easily offset by the reduced filling factor of the RF coil. However, Bennati and Orlando (2019) and Liu et al. (2017) demonstrated that large solution-state DNP enhancement factors can be achieved at 95 GHz. Another promising avenue for high-field, solution-state DNP experiments is microfluidic structures in combination with strip line resonators, and several research groups are active in this area (Denysenkov et al., 2017; Denysenkov and Prisner, 2019; Kratt et al., 2010; Webb, 2018). Alternatively, overmoded photonic band gap structures can be used for solution-state DNP; however, RF filling factors are typically low for these devices (Nevzorov et al., 2018). Direct polarization through DNP at high magnetic fields (14.1 T) of large sample volumes (> 100 µL) without using a resonator is possible in unpolar solvents (Dubroca et al., 2019). However, the microwave-induced sample heating limits this method to studying samples in solvents that have only small ohmic losses (e.g., small molecules in organic solvents), but even then chemical shift referencing will be challenging.

Over the past two decades, these technical challenges have led to the development of different strategies to avoid microwave-induced sample heating such as the dissolution DNP (dDNP) experiment pioneered by Ardenkjær-Larsen et al. (2003) in which the sample is polarized at low magnetic fields prior to melting it and quickly transferring it to a high-field spectrometer for NMR acquisition. The method can be used for analytical chemistry (Chen et al., 2013; Chen and Hilty, 2015) but is typically used to generate polarized solutions for magnetic resonance imaging (MRI) experiments (Ardenkjaer-Larsen, 2019, 2016). Although dDNP experiments are one-shot experiments, which cannot be repeated using the same sample, more dimensional experiments can be performed using rapid NMR methods (Frydman and Blazina, 2007). A variation of the dDNP experiment is the temperature jump experiment, in which the sample is polarized prior to melting the sample in situ using a powerful laser. Once the liquid-state NMR spectrum is recorded, the laser is switched off, the sample freezes, and it can be repolarized (Joo et al., 2009, 2006; Sharma et al., 2015). Alternatively, the liquid sample can be polarized at low magnetic fields (e.g., 0.35 T) and either rapidly shuttled into a high-field NMR spectrometer (e.g., 14.2 T) (Krahn et al., 2010; Reese et al., 2008, 2009) or transferred to the high-field magnet using a pump (Dorn et al., 1989, 1988; Liu et al., 2019). However, the polarization decreases during the transfer time, compromising the sensitivity gain achieved by the Overhauser dynamic nuclear polarization (ODNP) process. In addition, polarizing at a low magnetic field and detecting the NMR signal at a higher magnetic fields results in a “Boltzmann factor penalty” (Fedotov et al., 2020). For example, polarizing the sample at 0.35 T (15 MHz proton frequency) and detecting in a benchtop NMR system operation at 1.88 T (80 MHz proton frequency) results in a penalty factor of > 5, and acquiring the NMR signal at the same field at which it was polarized will always result in the largest overall sensitivity gain compared to methods that polarize at a low field and detect the signal at high fields.

Here, we employ ODNP spectroscopy to enhance the signal intensity in a low-field NMR experiment. The Overhauser effect causes a polarization transfer from an electron spin to a nuclear spin when the electron spin transition is irradiated. The amount of polarization transferred (ε) is given by the classic equation (Hausser and Stehlik, 1968):

(1) ε = 1 - E = ξ f s γ S γ I ,

where E is the enhancement, ξ is the coupling factor, f is the leakage factor, s is the electron saturation factor, γS is the electron gyromagnetic ratio, and γI is the nuclear gyromagnetic ratio. The coupling factor can vary from −1 in the case of pure scalar coupling to +0.5 in the case of pure dipolar coupling. For protons and nitroxides in solution, the interaction is almost entirely dipolar, which yields a maximum possible enhancement of 330. Hausser and Stehlik (1968) showed already in their early work that the Overhauser DNP effect is most efficient at low magnetic field strengths. The coupling factor dramatically decreases at higher magnetic fields leading to lower enhancement factors with increased magnetic field strengths.

Solution-state ODNP spectroscopy at low magnetic field strengths is no new method and was extensively used in the 1960s to the 1970s (Hausser and Stehlik, 1968; Mueller-Warmuth et al., 1983). However, the method almost completely vanished with the push of magnetic resonance methods to ever higher magnetic fields but was resurrected in the early 2000s because of its potential to determine local hydration dynamics on surfaces (Armstrong and Han, 2007). Today, it is an active field of research and the theory, instrumentation and application of ODNP spectroscopy is constantly developing (Armstrong and Han, 2009; Doll et al., 2012; Franck, 2020; Franck et al., 2013; Franck and Han, 2019; Han et al., 2008; Keller et al., 2020). While the observed enhancements can be large at low fields, a major drawback of low-field NMR spectroscopy, with or without the addition of ODNP, is the decreased resolution. However, the technique remains attractive for applications in which some of the resolution can be sacrificed at the benefit of greatly simplified instrumentation.

In an earlier publication we demonstrated that high-resolution, solution-state ODNP-enhanced NMR spectra can be recorded at low magnetic fields (0.35 T, 14 MHz 1H Larmor frequency) (Keller et al., 2020). Performing ODNP spectroscopy at this field has the advantage that instrumentation is readily available from X-band (9.5 GHz) EPR spectroscopy. In addition, enhancement factors are typically large, because the ODNP effect scales favorably with decreasing magnetic fields (Hausser and Stehlik, 1968; Kucuk et al., 2015; Sezer, 2014). Here, we present for the first time, ODNP-enhanced two-dimensional (2D) high-resolution proton NMR spectra of small molecules recorded at a magnetic field strength of 0.35 T using a highly homogenous permanent magnet. While 19F 2D ODNP-enhanced spectroscopy has have been reported previously, the small chemical shift dispersion of protons makes these experiments especially challenging (George and Chandrakumar, 2014). At a higher field of 1.2 T, ODNP experiments with 2D heteronuclear correlation (HETCOR) have been performed (Dey et al., 2017).

Experiments in this work were performed on a compact, home-built DNP/NMR system using a permanent magnet. Steps have been taken to compensate for temperature-induced magnetic field drift of the permanent magnet, which makes these experiments difficult. To mitigate these adverse effects and to obtain high-resolution spectra, we introduce a novel acquisition scheme and processing workflow.

2 Material and methods

2.1 Chemicals

4-Oxo-2,2,6,6-tetramethyl-1-piperidinyloxy (TEMPONE), 4-hydroxy-2,2,6,6-tetramethylpiperidine 1-oxyl (TEMPOL), ethyl crotonate, and ethanol were purchased from Sigma-Aldrich. All chemicals were used without further purification.

2.2 Sample preparation

For 10 mM TEMPONE in ethyl crotonate, 5 mm sample height was loaded into 0.98 mm ID, 1.00 mm OD quartz capillary (Hampton Research, HR6-146). For 10 mM TEMPOL in ethanol, 5 mm sample height was loaded into 0.60 mm ID, 0.84 mm OD quartz capillary (VitroCom, CV6084-Q-100).

2.3 ODNP spectrometer

All ODNP experiments were performed in a home-built spectrometer, which requires four principal components: (1) a high-power microwave source, (2) a microwave resonator with integrated NMR coil, (3) an NMR spectrometer, and (4) a magnet. We used a home-built microwave source with a maximum output power of 10 W, which operates over a frequency range of 8 to 12 GHz. A home-built, dielectric resonator operating in the TE011 mode at a frequency of 9.75 GHz with integrated saddle coil was used in all experiments, with a loaded quality factor Q of 6900. The Q dropped to > 4000 when a (aqueous) sample was inserted. The resonator is coupled to the waveguide by means of a circular iris. To optimize coupling to the resonator, the reflected power from the cavity was monitored and minimized by adjusting an iris screw. We estimated about 1.5 dB loss from the output of the microwave source to the sample position. Power levels given throughout this article correspond to the estimated microwave power levels at the position of the sample. The microwave source constantly monitors the forward (Tx) and reflected (Rx) microwave power. During operation the resonator can heat up, resulting in a shift of the resonator frequency, and the microwave frequency is automatically adjusted to minimize the reflected microwave power (lock mode).

NMR experiments were performed using a Kea2 spectrometer (Magritek), with an external RF amplifier (MiniCircuits, model LZY-22+). At the nominal output power of the RF amplifier of > 30 W, the observed NMR pulse length for a 90 pulse was about 5 µs.

All experiments are performed using a permanent dipole magnet (SABER Enterprises, LLC., North Andover, MA, USA). The nominal field strength of the magnet is 0.35 T, with a native homogeneity at the position of the sample of < 10 ppm. The magnet is equipped with a B0 sweep coil to make small adjustments (±15 mT) to the field strength. To perform high-resolution NMR spectroscopy, a set of electric shims are used. Shim coils were fabricated from printed circuit boards mounted to the magnet pole faces and included the zonal correction coils Z1 and Z2 and the tesseral correction coils X and Y. The physical dimensions of the coils were determined following the procedure outlined by Anderson (1961). Two triple channel power supplies (HP Model 6623A) were used to drive the sweep and shim coils. We observed a native linewidth of a (tap) water sample without energizing the shim coils of 110 Hz (8 ppm). With shim coils energized and optimized, we observed a linewidth of < 2.3 Hz (0.16 ppm) for a water sample with 200 µM TEMPOL (see Fig. S2 in the Supplement). Typically, the current for the electrical shims was optimized before each experiment. To compensate for ambient temperature fluctuations, the magnet was placed inside a small lab incubator. The temperature of the incubator was set to 32 C for all experiments. A picture of the experimental setup is shown in Fig. S1. To cool the sample, dry air was continuously flowed through the resonator at a rate of 2 L min−1.

2.4 ODNP experiments

The microwave power for all ODNP experiments was set to 35 dBm (3.2 W). The native resonance frequency of the ODNP resonator was found to be 9.75 GHz, corresponding to a proton NMR frequency of 14.7945 MHz. Continuous wave (cw) microwave irradiation was used for all experiments. The sweep coils of the magnet were set to maximize the ODNP enhancement.

One-dimensional proton ODNP spectra were acquired using an in-house-developed pulse program which allows for each phase cycle/average to be stored individually along a second dimension. One-dimensional proton ODNP spectra were acquired using a repetition time of 2 s and a total of 128 transients using a four-step phase cycle. Four dummy scans were performed before each acquisition to establish thermal equilibrium. The free induction decay (FID) contained 8192 points with a dwell time of 200 µs. A 90-pulse length of 5 µs was used.

J-resolved (JRES) experiments were acquired using an in-house-developed pulse program to save each phase cycle separately. A 1D proton reference spectrum was acquired after each phase cycle of the JRES experiment was completed (interleaved spectral referencing). Prior to each spectrum recorded in the t1 dimension, two dummy scans were used to equilibrate the magnetization during the experiment. The repetition time was set to 2.5 s, and a four-step phase cycle was used to eliminate artifacts (Berger and Braun, 2004). The spin echo was acquired with 4096 points and a dwell time of 200 µs. The indirect dimension was acquired with 128 points and a delay increment of 8 ms (initial inter-pulse delay of 3 ms). The 90- and 180-pulse lengths were set to 10 µs.

2.5 Data processing

All spectra were processed using DNPLab, an open-source Python package for processing DNP data (https://github.com/DNPLab/DNPLab, last access: 9 March 2021). The package is developed in collaboration between Bridge12 Technologies, Songi Han's lab at UCSB, and John Franck's lab at Syracuse University. DNPLab is able to import various spectrometer formats (e.g., Topspin, (Open) VnmrJ, Prospa, Tecmag) and converts the data into a versatile Python class for manipulating N-dimensional data arrays. Standard processing functions for NMR data can be easily applied along any specified dimension. In addition, DNPLab is flexible enough to allow experienced Python users to perform custom processing if desired. A complete description of DNPLab will be subject to a forthcoming publication.

To process 1D proton spectra, a window function was applied, prior to Fourier transformation of the FID. Averages were aligned using a FT cross-correlation method (Vu and Laukens, 2013) and summed together to generate the final spectrum.

The JRES data were corrected for field drift. A detailed description is given in Sect. 3.1. Typically, a Lorentz–Gauss transformation was applied along the direct and indirect dimension. Prior to Fourier transformation, the data were zero filled to twice the original length in both dimensions. After Fourier transformation, a shearing transformation was applied and the JRES spectrum was symmetrized using the geometric average (Ernst et al., 1987). The skyline projection was acquired by taking the maximum signal intensity along the indirect dimension. All ODNP-enhanced spectra, which have negative enhancements, are phased positively. No internal referencing standard was used. NMR spectra of ethyl crotonate were referenced according to values of the protons of the methyl group (at 1.28 ppm) as given by Berger and Braun (2004). NMR spectra of ethanol and water were referenced according to values given in Fulmer et al. (2010).

All experimental raw data are available in GitHub (DOI: https://doi.org/10.5281/zenodo.4479048) (Keller and Maly, 2021).

3 Results and discussion

To demonstrate the feasibility of low-field ODNP-enhanced 2D NMR spectroscopy, we choose two small molecules: ethanol and ethyl crotonate. While the NMR spectrum of ethanol is relatively simple (see Supplement for details), our discussion will focus on the use of ethyl crotonate. The NMR spectrum of ethyl crotonate is well-understood. The molecule has a variety of different proton sites with a large dispersion of proton chemical shifts and J-couplings, resulting in a “crowded” low-field NMR spectrum. The molecular structure for ethyl crotonate is given in Fig. 1.

https://mr.copernicus.org/articles/2/117/2021/mr-2-117-2021-f01

Figure 1Molecular structure of ethyl crotonate and proton labels used throughout this work.

Download

For the ethyl crotonate sample under study, we typically observe a signal enhancement of about 30, which is on the lower side. We attribute this to two factors: (1) the viscosity and (2) the sample temperature. In our studies we used a sample of 10 mM TEMPONE in neat ethyl crotonate. This sample has a higher viscosity than water, which will lead to lower overall enhancements (Hausser and Stehlik, 1968). Another factor that strongly influences the achieved enhancement is the sample temperature. Higher sample temperatures will lead to higher enhancements. However, due to the large resonator Q factor and the good separation of the microwave-induced electric and magnetic fields, sample heating is strongly minimized (Keller et al., 2020). However, we would like to note that even a moderate enhancement of about 30 will lead to a time saving factor of 900.

3.1 Interleaved spectral referencing for magnetic field drift correction

In general, magnetic field drifts cause line broadening and/or artifacts in NMR spectra, and numerous methods have been implemented for solution and solid-state NMR experiments to compensate for these adverse effects. These hardware-based methods typically use a lock signal from a reference sample or nucleus (e.g., deuterium), and either the field or the NMR transmitter frequency is adjusted accordingly (Baker and Burd, 1957; Maly et al., 2006; Markiewicz, 2002; Paulson and Zilm, 2005). In contrast, software-based methods can be also used to account for magnetic field drifts (Ha et al., 2014; Najbauer and Andreas, 2019).

Using a software-based method to correct for the field drift has the advantage that no additional hardware is required. We used a compact, single-channel NMR spectrometer with no lock channel. This type of hardware is commonly found in NMR applications for process monitoring or well-logging applications, which often have strict space restrictions, and adding additional hardware is challenging or simply not possible (Ha et al., 2014; Mandal et al., 2014; Song and Kausik, 2019).

Correcting for magnetic field drift can be particularly challenging for permanent magnets utilizing rare-earth magnets which have a large temperature coefficient causing the field to be susceptible to small fluctuations in room temperature. In such magnets it is common to stabilize the temperature to  100 mK (Windt et al., 2011) in addition to using a field-frequency lock (Blümich, 2016; Danieli et al., 2010).

The temperature drift will cause a slow drift over the course of minutes or hours. Higher-frequency fluctuations caused by magnetic coupling to the environment will also lead to line broadening and/or artifacts. Ripple from shim power supplies can cause line broadening and/or sidebands at the mains frequency. In addition, proportional–integral–derivative (PID) temperature controls which typically use pulse width modulation (PWM) to control heaters can cause field shifts and/or oscillations.

https://mr.copernicus.org/articles/2/117/2021/mr-2-117-2021-f02

Figure 2(a) One-dimensional ODNP-enhanced ethyl crotonate reference spectra with 10 mM TEMPONE acquired after each completed phase cycle step of JRES experiment. Spectra are vertically offset for clarity. (b) Chemical shift of the peak with the maximum intensity and fit (fourth-order polynomial) for experimental drift correction. (c) Residual of the fit.

Download

In our experiments, we employ two strategies to minimize artifacts from field drift. Any high-frequency fluctuations were suppressed by disabling the heater during NMR acquisition. In this case, the temperature will no longer be regulated, and the field will start to drift. To correct this, we acquired a reference 1D proton NMR reference spectrum after each complete phase cycle of the JRES experiment (interleaved spectral referencing; see Fig. 2a). Since we only record a reference spectrum after each phase cycle is completed, the overall measurement time is only slightly increased, adding only little overhead. For the JRES experiments, we used a four-step phase cycle; therefore, interleaved spectral referencing only adds 20 % overhead. If the drift is small, fewer reference spectra can be recorded, further reducing the amount of overhead.

Over the period of acquiring the 2D JRES spectrum we typically observe an overall field drift of approximately 60 ppm over the course of  30 min. The change in chemical shift was calculated relative to the first spectrum, and the change in chemical shift for each JRES step was then fit to a fourth-order polynomial Fig. 2b. We found a fourth-order polynomial to be sufficient, resulting in a residual <±0.05 ppm (see Fig. 2c), significantly smaller compared to the observed proton linewidth of  3 Hz (0.2 ppm). Each transient of the JRES experiment, including different phase cycle steps, was stored separately so that a field drift correction could be applied before the phase cycle. In addition to chemical shift, the 1D proton reference spectra were also used to correct the spectral phase of each step in the JRES experiment (a detailed description is given in the Supplement). In addition to the field drift, we also observed a phase drift over the course of the experiment, which we attribute to two different effects: (1) the acquisition delay (spectrometer dead time) will result in a linear phase roll in the frequency domain as the field drifts, and there are (2) instrumental and environmental instabilities (e.g., ambient temperature fluctuations in the room). All these effects can be corrected for by interleaved spectral referencing. All processing was performed using the Python package DNPLab.

Interleaved spectra referencing is a general concept, which is not restricted to any particular NMR sequence. The frequency of the acquisition of the reference spectrum depends on the scale of the observed field drift, only adding very little overhead to the entire acquisition. Acquiring a separate 1D spectrum has the additional benefit that the signal intensity will be always constant. In a 2D experiment transients with longer evolution times often do not have a sufficiently large enough signal-to-noise ratio. In the two presented cases for ethyl crotonate and ethanol, we only observed a monotonic drift of the field. However, as long as a function can be found that adequately models the field drift it will be possible to correct for the drift even for non-monotonic drifts (e.g., oscillations) or sudden field jumps.

3.2 Resolution enhancement

In Fig. 3 (top), a 1D proton NMR spectrum of ethyl crotonate is shown together with the resonance assignment corresponding to the protons of the molecule. All resonances can be identified and assigned. We applied a 1 Hz Lorentzian apodization window resulting in an average native linewidth of 7.3 Hz (see the Supplement for details). As demonstrated in an earlier publication, the presence of the polarizing agent (TEMPONE) does not prevent us from resolving the J-couplings in ethyl crotonate (Keller et al., 2020). However, the polarizing agent does introduce significant line broadening at a concentration of 10 mM, and the linewidth is limited by the polarizing agent, not magnetic field homogeneity. The linewidth limited by the magnetic field is 2.3 Hz (0.16 ppm) as shown for a sample of 200 µM TEMPOL in water (see Supplement, Fig. S2).

https://mr.copernicus.org/articles/2/117/2021/mr-2-117-2021-f03

Figure 3One-dimensional ODNP-enhanced NMR spectra (128 averages) of ethyl crotonate with 10 mM TEMPONE using different apodization functions. Top: 1 Hz Lorentzian line broadening. Middle: Lorentz–Gauss transformation using 4 Hz line broadening for both Gaussian and Lorentzian linewidths. Bottom: resolution enhancement after Traficante and Nemeth (1987) and Traficante and Rajabzadeh (2000) using T2 of 0.2 s.

Download

The spectral resolution can be further improved using common resolution enhancement techniques such as Lorentz–Gauss transformation or other windowing functions (Ernst et al., 1987; Traficante and Nemeth, 1987; Traficante and Rajabzadeh, 2000). Typically, these methods are known to decrease the signal-to-noise ratio; however, because of the dramatic improvement in sensitivity when using ODNP, some loss in spectral resolution for low-field NMR experiments can be regained. This is especially attractive, since the observed ODNP enhancements are larger at lower fields. The results for two commonly used window functions for resolution enhancement are shown in Fig. 3 (middle and bottom). By using the Lorentz–Gauss transformation, the overall spectral linewidth can be reduced from 7.3 Hz, as observed in the NMR spectrum process using a 1 Hz Lorentzian line broadening, to 5.6 Hz (Fig. 3, middle). Further reduction in linewidth can be achieved using the method reported by Traficante and Nemeth (1987) and Traficante and Rajabzadeh (2000), however, at the expense of some increased artifacts (Fig. 3, bottom). Using this window function a linewidth of 4.4 Hz was observed. Both the Lorentz–Gauss transformation and the resolution-enhancement window function improve the observed linewidth. However, for the 2D JRES experiments we used the Lorentz–Gauss transformation, since it introduces fewer artifacts. For a detailed discussion about the choice of the apodization window, the reader is referred to the discussion in the Supplement.

3.3 ODNP-enhanced 1H JRES experiments on ethyl crotonate

At low magnetic fields, overlapping J-couplings can lead to crowded, unresolved spectra. One method to simplify complicated spectra is JRES spectroscopy, which separates the J-coupling along a second dimension (Aue et al., 1976). This technique is commonly used in metabolomics (Ludwig and Viant, 2010). The JRES experiment is the simplest experiment of methods belonging to a group of experiments commonly referred to as pure-shift experiments (Aguilar et al., 2010; Zangger, 2015). We used the JRES method, the simplest implementation of pure-shift spectroscopy, because it does not require the use of pulsed field gradients.

https://mr.copernicus.org/articles/2/117/2021/mr-2-117-2021-f04

Figure 4(a) ODNP-enhanced 1H JRES spectrum of neat ethyl crotonate with 10 mM TEMPONE. The dashed boxes indicate areas showing strong coupling effects. (b) Slices of JRES spectrum shown in panel (a) for each proton in ethyl crotonate. The expected multiplet pattern is indicated by vertical lines with corresponding tick marks. Spectra are offset for clarity. Features marked with an asterisk are artifacts of the JRES experiment.

Download

JRES experiments were performed on samples of two small molecules, ethanol and ethyl crotonate. Due to the limited features observed in the JRES experiment, the data for ethanol are shown in the Supplement. The ODNP-enhanced JRES spectrum of ethyl crotonate with 10 mM TEMPONE is shown in Fig. 4a. For each chemical shift corresponding to a specific proton site, a 1D slice is given showing the pattern caused by the J-coupling for each proton site (Fig. 4b, labels correspond to protons as shown in Fig. 1). The protons of the two methyl groups (protons A and B) which are overlapping in the 1D NMR spectrum (Fig. 3) can be clearly resolved in the 2D JRES spectrum. The same is true for the methylene quartet (proton C) which has some overlap with proton D in the 1D spectrum. The measured J-couplings are 7 ± 1 Hz for J(H-A, H-C), 7 ± 1 Hz for J(H-B, H-E), and 16 ± 1 Hz for J(H-D, H-E). The observed values are in excellent agreement to within 1 Hz with values published in the literature (Berger and Braun, 2004).

To obtain a 1D spectrum, in which the J-coupling is removed (pure-shift spectrum), a skyline projection of the JRES spectrum along the J-coupling dimension can be calculated (see Fig. 5, orange). While this simplifies the spectrum, the peak intensities are not quantitative anymore. For a quantitative analysis, the peak intensities have to be integrated along the indirect dimension, which will allow for accurate measurements of the enhancements for each chemical site (see Supplement). In the skyline projection, the peaks corresponding to the methyl groups (protons A and B) are clearly separated, compared to a regular 1D proton spectrum (Fig. 5, green). In addition, the quartet located at 4.2 ppm in the regular 1D spectrum collapses to a single, well-resolved line in the skyline projection located at a chemical shift of 4.2 ppm.

https://mr.copernicus.org/articles/2/117/2021/mr-2-117-2021-f05

Figure 5NMR spectra of ethyl crotonate. Top: skyline projection of 2D JRES experiment. The locations of strong-coupling artifacts are indicated by asterisks. Bottom: 1D 1H NMR spectrum.

Download

It is a well-known fact that JRES spectra suffer from artifacts resulting from strongly coupled spin systems. These artifacts cannot be removed by phase cycling or pulsed field gradients, since these artifacts originate from the same coherence transfer path as the desired signal. These artifacts lead to additional lines in the skyline projection which are not at the position of the chemical shift (Ludwig and Viant, 2010; Thrippleton et al., 2005). For ethyl crotonate, we observe strong-coupling artifacts for protons A and C at a chemical shift of about 3 ppm and a J-coupling of ±20 Hz in the ODNP-enhanced JRES spectrum. Furthermore, peaks due to strong coupling effects can be observed in the region of 6 to 7 ppm. These peaks are attributed to the coupling between protons D and E and result in a complicated multiplet structure (see dashed boxes in Fig. 4a). These unwanted peaks also show up in the skyline projection shown in Fig. 5 (orange, marked with an asterisk). However, they are easy to distinguish in the full 2D spectrum, and since the skyline projection is calculated from the 2D spectrum, it is relatively straightforward to separate wanted from unwanted peaks.

4 Future directions

Currently, the majority of low-field, ODNP-enhanced NMR experiments are 1D experiments to study hydration dynamics. For these measurements a decent homogeneity of the magnetic field is sufficient since only a single peak (water) is observed. However, improving the homogeneity of the magnetic field will not only lead to higher sensitivity due to improved line shapes, but it also opens the possibilities of performing high-resolution ODNP-enhanced NMR spectroscopy. The use of active shims can improve spectral resolution to the linewidth limited by the paramagnetic broadening effects of the polarizing agent. Any further improvement in resolution must be gained by other techniques such as resolution enhancement (data processing) and/or multi-dimensional NMR experiments, and the possible applications are plentiful.

As a new application, high-resolution ODNP-enhanced JRES spectroscopy enables site-specific, chemical shift-resolved solvent dynamics measurements. For example, ODNP measurements on toluene show different saturation behavior for the methyl and aromatic protons in toluene, suggesting different solvent dynamics for the different proton sites (Enkin et al., 2014; Keller et al., 2020). Since the peak separation is only about 4 ppm, highly homogenous magnetic fields are required to study the peaks individually since overlapping peaks will make data analysis more difficult. These types of experiments can greatly benefit from JRES spectroscopy. In contrast to correlation type experiments such as COSY, the individual proton sites are resolved in the indirect dimension in the JRES experiment. This will allow detecting the microwave power saturation behavior of an individual proton site. A correlation peak observed in for example a COSY experiment on the other hand will show the superposition of the saturation behavior of two sites, further complicating the data analysis.

Another recent example is the study of water diffusion in polymer membranes such as Nafion. Water located inside the membrane (channel water) has a higher chemical shift of about 5.5 ppm, while residual water has a chemical shift of 4.7 ppm (Kim et al., 2016; Überrück et al., 2018). To study the diffusion behavior using ODNP spectroscopy high-resolution techniques are required in combination with multidimensional NMR methods. Two-dimensional ODNP-enhanced JRES experiments are particularly helpful because they allow studying complex systems by bringing site-specific resolution.

A common practice in low-field NMR spectroscopy is to record T1/T2 relaxation maps to identify different species in complex mixtures based on their relaxation behavior (Colnago et al., 2021; Song et al., 2021). Other methods just rely on recording either T1 or T2 relaxation times and are successfully used in cancer diagnosis (Castro et al., 2014; Issadore et al., 2011; Min et al., 2012). These point-of-care systems can potentially benefit from ODNP-enhanced, multi-dimensional, low-field NMR spectroscopy, not only by studying the relaxation behavior of a sample but also by acquiring spectral information.

Another field that will benefit from the presented technique is NMR spectroscopy used in reaction monitoring (Dalitz et al., 2012; Plainchont et al., 2018). High-resolution ODNP-enhanced 2D NMR spectroscopy opens the possibility of studying more complex mixtures. Portable low-field ODNP systems have been reported in the literature (Ebert et al., 2012; Münnemann et al., 2008), but so far these systems only achieve low to moderate spectral resolution. Typically, compact, portable systems are used in these applications that do not use a superconducting magnet. These systems are based on small electromagnets or permanent magnets, which can greatly benefit not only from ODNP spectroscopy but also from interleaved spectral referencing.

From here the possibilities of improvements are countless. Implementation of JRES methods that do not result in artifacts due to strong spin coupling effects is straightforward (Thrippleton et al., 2005). Furthermore, integration of pulse field gradient coils into the magnet system will accelerate acquisition of spectral acquisition of multidimensional experiments. In addition, with strong enough gradient pulses, ultrafast 2D methods introduced by Frydman and Blazina (2007) are possible (Gouilleux et al., 2018). Many of these concepts are currently implemented in our lab.

5 Conclusion

In this work, we demonstrate the application of ODNP-enhanced 2D JRES spectroscopy to improve spectral resolution beyond the limit given by the line broadening introduced by the paramagnetic polarizing agent. As a proof of concept we use the simplest implementation of the 2D JRES experiment and achieve full spectral resolution for small molecules at low magnetic fields.

We show that multi-dimensional NMR experiments can be applied to increase the resolution in low-field ODNP experiments. Using ODNP-enhanced 2D JRES spectroscopy, we are able to separate the overlapping multiplets of ethyl crotonate into a second dimension and clearly identify each chemical cite. While in most circumstances the emphasis of (O)DNP-enhanced spectroscopy is on improvement in sensitivity, it should be noted that the improved sensitivity can allow more aggressive apodization of data to increase spectral resolution.

Crucial to these experiments is the interleaved spectral referencing to compensate for temperature-induced field drifts over the course of the JRES experiment. This method does not require additional hardware such as a field-frequency lock, which is especially challenging when designing compact systems.

Code availability

All experimental data were processed using DNPLab, an open-source Python package to process NMR, EPR, and DNP data. DNPLab is available for free download at https://github.com/DNPLab/DNPLab (https://doi.org/10.5281/zenodo.4670553, Keller et al., 2021).

Data availability

Experimental raw data are publicly available on GitHub at https://github.com/Bridge12Technologies/2D_ODNP_Spectroscopy_DataRep (last access: 29 January 2021; https://doi.org/10.5281/zenodo.4479048, Keller and Maly, 2021).

Supplement

The supplement related to this article is available online at: https://doi.org/10.5194/mr-2-117-2021-supplement.

Author contributions

TM designed the research and designed the experiments in collaboration with TJK. All experiments were carried out and analyzed by TJK.

Competing interests

Timothy J. Keller and Thorsten Maly are employees of Bridge12 Technologies, Inc. Thorsten Maly is a co-founder of Bridge12 Technologies, Inc.

Acknowledgements

We thank Dennis Gautreau, Walter Hrynyk, Charan Gujjala, and Rohit Arora for the mechanical design and John Franck (Syracuse University), Thomas Casey (University of California, Santa Barbara), Songi Han (University of California, Santa Barbara), Jagadishwar Sirigiri, and Alexander Laut for many stimulating discussions.

Financial support

This research has been supported by the National Institute of General Medical Sciences (grant nos. GM112391 and GM128542).

Review statement

This paper was edited by Patrick Giraudeau and reviewed by Gunnar Jeschke and one anonymous referee.

References

Aguilar, J. A., Faulkner, S., Nilsson, M., and Morris, G. A.: Pure Shift 1H NMR: A Resolution of the Resolution Problem?, Angew. Chem. Int. Ed., 49, 3901–3903, https://doi.org/10.1002/anie.201001107, 2010. 

Anderson, W. A.: Electrical Current Shims for Correcting Magnetic Fields, Rev. Sci. Instrum., 32, 241–250, https://doi.org/10.1063/1.1717338, 1961. 

Ardenkjaer-Larsen, J. H.: On the present and future of dissolution-DNP, J. Magn. Reson., 264, 3–12, https://doi.org/10.1016/j.jmr.2016.01.015, 2016. 

Ardenkjaer-Larsen, J. H.: Hyperpolarized MR – What's up Doc?, J. Magn. Reson., 306, 124–127, https://doi.org/10.1016/j.jmr.2019.07.017, 2019. 

Ardenkjær-Larsen, J. H., Fridlund, B., Gram, A., Hansson, G., Hansson, L., Lerche, M. H., Servin, R., Thaning, M., and Golman, K.: Increase in signal-to-noise ratio of > 10 000 times in liquid-state NMR, Proc. Natl. Acad. Sci., 100, 10158–10163, https://doi.org/10.1073/pnas.1733835100, 2003. 

Armstrong, B. D. and Han, S.: A new model for Overhauser enhanced nuclear magnetic resonance using nitroxide radicals, J. Chem. Phys., 127, 104508, https://doi.org/10.1063/1.2770465, 2007. 

Armstrong, B. D. and Han, S.: Overhauser Dynamic Nuclear Polarization To Study Local Water Dynamics, J. Am. Chem. Soc., 131, 4641–4647, https://doi.org/10.1021/ja809259q, 2009. 

Aue, W. P., Karhan, J., and Ernst, R. R.: Homonuclear broad band decoupling and two-dimensional J-resolved NMR spectroscopy, J. Chem. Phys., 64, 4226–4227, https://doi.org/10.1063/1.431994, 1976. 

Baker, E. B. and Burd, L. W.: High Stability Nuclear Magnetic Resonance Spectrograph, Rev. Sci. Instrum., 28, 313–321, https://doi.org/10.1063/1.1715873, 1957. 

Bennati, M. and Orlando, T.: Overhauser DNP in Liquids on 13C Nuclei, 8, emrstm1581, https://doi.org/10.1002/9780470034590, 2019. 

Berger, S. and Braun, S.: 200 and More NMR Experiments: A Practical Course, Wiley-VCH, Weinheim, Germany, 854 pp., 2004. 

Blümich, B.: Introduction to compact NMR: A review of methods, TrAC Trends Anal. Chem., 83, 2–11, https://doi.org/10.1016/j.trac.2015.12.012, 2016. 

Castro, C. M., Ghazani, A. A., Chung, J., Shao, H., Issadore, D., Yoon, T.-J., Weissleder, R., and Lee, H.: Miniaturized nuclear magnetic resonance platform for detection and profiling of circulating tumor cells, Lab. Chip, 14, 14–23, https://doi.org/10.1039/C3LC50621E, 2014. 

Chen, H. Y. and Hilty, C.: Implementation and characterization of flow injection in dissolution dynamic nuclear polarization NMR spectroscopy, Chemphyschem, 16, 2646–2652, https://doi.org/10.1002/cphc.201500292, 2015. 

Chen, H. Y., Ragavan, M., and Hilty, C.: Protein folding studied by dissolution dynamic nuclear polarization, Angew. Chem., 52, 9192–9195, https://doi.org/10.1002/anie.201301851, 2013. 

Colnago, L. A., Wiesman, Z., Pages, G., Musse, M., Monaretto, T., Windt, C. W., and Rondeau-Mouro, C.: Low field, time domain NMR in the agriculture and agrifood sectors: An overview of applications in plants, foods and biofuels, J. Magn. Reson., 323, 106899, https://doi.org/10.1016/j.jmr.2020.106899, 2021. 

Corzilius, B.: Paramagnetic Metal Ions for Dynamic Nuclear Polarization, EMagRes, 7, 179–194, 2018. 

Dalitz, F., Cudaj, M., Maiwald, M., and Guthausen, G.: Process and reaction monitoring by low-field NMR spectroscopy, Prog. Nucl. Magn. Reson. Spectrosc., 60, 52–70, https://doi.org/10.1016/j.pnmrs.2011.11.003, 2012. 

Danieli, E., Perlo, J., Blümich, B., and Casanova, F.: Small Magnets for Portable NMR Spectrometers, Angew. Chem. Int. Ed., 49, 4133–4135, https://doi.org/10.1002/anie.201000221, 2010. 

Denysenkov, V. P. and Prisner, T. F.: Liquid-state Overhauser DNP at High Magnetic Fields, EMagRes, 8, 41–54, https://onlinelibrary.wiley.com/doi/abs/10.1002/9780470034590.emrstm1557 (last access: 8 April 2021), 2019. 

Denysenkov, V. P., Terekhov, M., Maeder, R., Fischer, S., Zangos, S., Vogl, T., and Prisner, T. F.: Continuous-flow DNP polarizer for MRI applications at 1.5 T, Sci. Rep., 7, 44010, https://doi.org/10.1038/srep44010, 2017. 

Dey, A., Banerjee, A., and Chandrakumar, N.: Transferred Overhauser DNP: A Fast, Efficient Approach for Room Temperature 13C ODNP at Moderately Low Fields and Natural Abundance, J. Phys. Chem. B, 121, 7156–7162, https://doi.org/10.1021/acs.jpcb.7b05081, 2017. 

Doll, A., Bordignon, E., Joseph, B., Tschaggelar, R., and Jeschke, G.: Liquid state DNP for water accessibility measurements on spin-labeled membrane proteins at physiological temperatures, J. Magn. Reson., 222, 34–43, https://doi.org/10.1016/j.jmr.2012.06.003, 2012. 

Dorn, H. C., Wang, J., Allen, L., Sweeney, D., and Glass, T. E.: Flow dynamic nuclear polarization, a novel method for enhancing NMR signals in flowing fluids, J. Magn. Reson., 79, 404–412, https://doi.org/10.1016/0022-2364(88)90078-9, 1988. 

Dorn, H. C., Gitti, R., Tsai, K. H., and Glass, T. E.: The flow transfer of a bolus with 1H dynamic nuclear polarization from low to high magnetic fields, Chem. Phys. Lett., 155, 227–232, https://doi.org/10.1016/0009-2614(89)85354-0, 1989. 

Dubroca, T., Wi, S., van Tol, J., Frydman, L., and Hill, S.: Large volume liquid state scalar Overhauser dynamic nuclear polarization at high magnetic field, Phys. Chem. Chem. Phys., 21, 21200–21204, https://doi.org/10.1039/C9CP02997D, 2019. 

Ebert, S., Amar, A., Bauer, C., Kölzer, M., Blümler, P., Spiess, H., Hinderberger, D., and Münnemann, K.: A Mobile DNP Polarizer for Continuous Flow Applications, Appl. Magn. Reson., 43, 1–12, https://doi.org/10.1007/s00723-012-0344-7, 2012. 

Enkin, N., Liu, G., Tkach, I., and Bennati, M.: High DNP efficiency of TEMPONE radicals in liquid toluene at low concentrations, Phys. Chem. Chem. Phys. PCCP, 16, 8795–8800, https://doi.org/10.1039/c4cp00854e, 2014. 

Ernst, R. R., Bodenhausen, G., and Wokaun, A.: Principles of Nuclear Magnetic Resonance in One and Two Dimensions, International Series of Monographs on Chemistry 14, Oxford University Press, Oxford, UK, 1987. 

Fedotov, A., Kurakin, I., Fischer, S., Vogl, T., Prisner, T. F., and Denysenkov, V.: Increased flow rate of hyperpolarized aqueous solution for dynamic nuclear polarization-enhanced magnetic resonance imaging achieved by an open Fabry–Pérot type microwave resonator, Magn. Reson., 1, 275–284, https://doi.org/10.5194/mr-1-275-2020, 2020. 

Franck, J. M.: Overhauser Dynamic Nuclear Polarization: A Tool for Building Maps of Hydration Water, Biophys. J., 118, 487a, https://doi.org/10.1016/j.bpj.2019.11.2695, 2020. 

Franck, J. M. and Han, S.: Overhauser Dynamic Nuclear Polarization for the Study of Hydration Dynamics, Explained, Methods Enzymol., 615, 131–175, https://doi.org/10.1016/bs.mie.2018.09.024, 2019. 

Franck, J. M., Pavlova, A., Scott, J. A., and Han, S.: Quantitative cw Overhauser effect dynamic nuclear polarization for the analysis of local water dynamics, Prog. Nucl. Magn. Reson. Spectrosc., 74, 33–56, https://doi.org/10.1016/j.pnmrs.2013.06.001, 2013. 

Frydman, L. and Blazina, D.: Ultrafast two-dimensional nuclear magnetic resonance spectroscopy of hyperpolarized solutions, Nat. Phys., 3, 415–419, 2007. 

Fulmer, G. R., Miller, A. J. M., Sherden, N. H., Gottlieb, H. E., Nudelman, A., Stoltz, B. M., Bercaw, J. E., and Goldberg, K. I.: NMR Chemical Shifts of Trace Impurities: Common Laboratory Solvents, Organics, and Gases in Deuterated Solvents Relevant to the Organometallic Chemist, Organometallics, 29, 2176–2179, https://doi.org/10.1021/om100106e, 2010. 

George, C. and Chandrakumar, N.: Chemical-Shift-Resolved19F NMR Spectroscopy between 13.5 and 135 MHz: Overhauser-DNP-Enhanced Diagonal Suppressed Correlation Spectroscopy, Angew. Chem., 126, 8581–8584, https://doi.org/10.1002/ange.201402320, 2014. 

Gouilleux, B., Rouger, L., and Giraudeau, P.: Ultrafast 2D NMR: Methods and Applications, Annu. Rep. NMR Spectro., 93, 75–144, https://doi.org/10.1016/bs.arnmr.2017.08.003, 2018. 

Griffin, R. G., Swager, T. M., and Temkin, R. J.: High frequency dynamic nuclear polarization: New directions for the 21st century, J. Magn. Reson., 306, 128–133, https://doi.org/10.1016/j.jmr.2019.07.019, 2019. 

Ha, D., Paulsen, J., Sun, N., Song, Y.-Q., and Ham, D.: Scalable NMR spectroscopy with semiconductor chips, Proc. Natl. Acad. Sci., 111, 11955–11960, https://doi.org/10.1073/pnas.1402015111, 2014. 

Han, S., McCarney, E. R., and Armstrong, B. D.: Dynamic Nuclear Polarization Studies of Local Water Dynamics in Soft Molecular Assemblies at 9.8 GHz, Appl. Magn. Reson., 34, 439–451, 2008. 

Hausser, K. H. and Stehlik, D.: Dynamic Nuclear Polarization in Liquids, Academic Press, New York, USA, 79–139, 1968. 

Issadore, D., Min, C., Liong, M., Chung, J., Weissleder, R., and Lee, H.: Miniature magnetic resonance system for point-of-care diagnostics, Lab Chip, 11, 2282–2287, https://doi.org/10.1039/c1lc20177h, 2011. 

Jaudzems, K., Polenova, T., Pintacuda, G., Oschkinat, H., and Lesage, A.: DNP NMR of biomolecular assemblies, J. Struct. Biol., 206, 90–98, https://doi.org/10.1016/j.jsb.2018.09.011, 2019. 

Joo, C. G., Hu, K. N., Bryant, J. A., and Griffin, R. G.: In situ temperature jump high-frequency dynamic nuclear polarization experiments: enhanced sensitivity in liquid-state NMR spectroscopy., J. Am. Chem. Soc., 128, 9428–32, 2006. 

Joo, C.-G., Casey, A., Turner, C. J., and Griffin, R. G.: In Situ Temperature-Jump Dynamic Nuclear Polarization: Enhanced Sensitivity in Two Dimensional 13C-13C Correlation Spectroscopy in Solution, J. Am. Chem. Soc., 131, 12–13, https://doi.org/10.1021/ja805521y, 2009. 

Kaminker, I.: Recent Advances in Magic Angle Spinning-Dynamic Nuclear Polarization Methodology, Isr. J. Chem., 59, 990–1000, https://doi.org/10.1002/ijch.201900092, 2019. 

Keller, T. J. and Maly, T.: Overhauser Dynamic Nuclear Polarization Enhanced Two-Dimensional Proton NMR Spectroscopy at Low Magnetic Fields, Zenodo, https://doi.org/10.5281/zenodo.4479048, 2021. 

Keller, T. J., Laut, A. J., Sirigiri, J., and Maly, T.: High-Resolution Overhauser Dynamic Nuclear Polarization Enhanced Proton NMR Spectroscopy at Low Magnetic Fields, J. Magn. Reson., 313, 106719, https://doi.org/10.1016/j.jmr.2020.106719, 2020. 

Keller, T., Casey, T., Lin, Y., Maly, T., and Franck, J. M.: DNPLab/DNPLab (Version 2DODNPPaper), Zenodo, https://doi.org/10.5281/zenodo.4670553, 2021. 

Kim, H. N., Hwang, R. Y., and Han, O. H.: Behavior of Channel Water and CF2H Side-Chain Terminal Groups in Swollen Nafion Polymer Electrolyte Membranes after Thermal Treatment, ACS Macro Lett., 5, 801–804, https://doi.org/10.1021/acsmacrolett.6b00388, 2016. 

Krahn, A., Lottmann, P., Marquardsen, T., Tavernier, A., Turke, M.-T., Reese, M., Leonov, A., Bennati, M., Hoefer, P., Engelke, F., and Griesinger, C.: Shuttle DNP spectrometer with a two-center magnet, Phys. Chem. Chem. Phys., 12, 5830–5840, https://doi.org/10.1039/C003381B, 2010. 

Kratt, K., Badilita, V., Burger, T., Korvink, J. G., and Wallrabe, U.: A fully MEMS-compatible process for 3D high aspect ratio micro coils obtained with an automatic wire bonder, J. Micromech. Microeng., 20, 015021, https://doi.org/10.1088/0960-1317/20/1/015021, 2010. 

Kucuk, S. E., Neugebauer, P., Prisner, T. F., and Sezer, D.: Molecular simulations for dynamic nuclear polarization in liquids: a case study of TEMPOL in acetone and DMSO, Phys. Chem. Chem. Phys., 17, 6618–6628, https://doi.org/10.1039/c4cp05832a, 2015. 

Liao, W.-C., Ghaffari, B., Gordon, C. P., Xu, J., and Copéret, C.: Dynamic Nuclear Polarization Surface Enhanced NMR spectroscopy (DNP SENS): Principles, protocols, and practice, Curr. Opin. Colloid Interface Sci., 33, 63–71, https://doi.org/10.1016/j.cocis.2018.02.006, 2018. 

Liu, G., Levien, M., Karschin, N., Parigi, G., Luchinat, C., and Bennati, M.: One-thousand-fold enhancement of high field liquid nuclear magnetic resonance signals at room temperature, Nat. Chem., 9, 676–680, https://doi.org/10.1038/nchem.2723, 2017. 

Liu, X., Gu, J., Wightman, J., and Dorn, H. C.: Elucidation of Oxygen Chemisorption Sites on Activated Carbons by 1H DNP for Insight into Oxygen Reduction Reactions, ACS Appl. Nano Mater., 2, acsanm.9b01308, https://doi.org/10.1021/acsanm.9b01308, 2019. 

Ludwig, C. and Viant, M. R.: Two-dimensional J-resolved NMR spectroscopy: review of a key methodology in the metabolomics toolbox, Phytochem. Anal., 21, 22–32, https://doi.org/10.1002/pca.1186, 2010. 

Maly, T., Bryant, J., Ruben, D., and Griffin, R. G.: A field-sweep/field-lock system for superconducting magnets – Application to high-field EPR, J. Magn. Reson., 183, 303–307, 2006. 

Maly, T., Debelouchina, G. T., Bajaj, V. S., Hu, K.-N., Joo, C.-G., MakJurkauskas, M. L., Sirigiri, J. R., van der Wel, P. C. A., Herzfeld, J., Temkin, R. J., and Griffin, R. G.: Dynamic nuclear polarization at high magnetic fields, J. Chem. Phys., 128, 052211, https://doi.org/10.1063/1.2833582, 2008. 

Mandal, S., Utsuzawa, S., Cory, D. G., Hürlimann, M., Poitzsch, M., and Song, Y. Q.: An ultra-broadband low-frequency magnetic resonance system, J. Magn. Reson., 242, 113–125, https://doi.org/10.1016/j.jmr.2014.02.019, 2014. 

Markiewicz, W. D.: Current injection for field decay compensation in NMR spectrometer magnets, IEEE Trans. Appl. Supercond., 12, 1886–1890, https://doi.org/10.1109/TASC.2002.806027, 2002. 

Min, C., Shao, H., Liong, M., Yoon, T.-J., Weissleder, R., and Lee, H.: Mechanism of Magnetic Relaxation Switching Sensing, ACS Nano, 6, 6821–6828, https://doi.org/10.1021/nn301615b, 2012. 

Mueller-Warmuth, W., Meise-Gresch, K., and Waugh, J. S.: Molecular Motions and Interactions as Studied by Dynamic Nuclear Polarization (DNP) in Free Radical Solutions, Academic Press, New York, USA, 1–45, 1983. 

Münnemann, K., Bauer, C., Schmiedeskamp, J., Spiess, H. W., Schreiber, W. G., and Hinderberger, D.: A Mobile DNP Polarizer for Clinical Applications, Appl. Magn. Reson., 34, 321–330, 2008. 

Najbauer, E. E. and Andreas, L. B.: Correcting for magnetic field drift in magic-angle spinning NMR datasets, J. Magn. Reson., 305, 1–4, https://doi.org/10.1016/j.jmr.2019.05.005, 2019. 

Nanni, E. A., Barnes, A. B., Matsuki, Y., Woskov, P. P., Corzilius, B., Griffin, R. G., and Temkin, R. J.: Microwave field distribution in a magic angle spinning dynamic nuclear polarization NMR probe, J. Magn. Reson., 210, 16–23, https://doi.org/10.1016/j.jmr.2011.02.001, 2011. 

Neumann, M.: The dielectric constant of water. Computer simulations with the MCY potential, J. Chem. Phys., 82, 5663–5672, https://doi.org/10.1063/1.448553, 1985. 

Nevzorov, A. A., Milikisiyants, S., Marek, A. N., and Smirnov, A. I.: Multi-resonant photonic band-gap/saddle coil DNP probehead for static solid state NMR of microliter volume samples, J. Magn. Reson., 297, 113–123, https://doi.org/10.1016/j.jmr.2018.10.010, 2018. 

Paulson, E. K. and Zilm, K. W.: External field-frequency lock probe for high resolution solid state NMR, Rev. Sci. Instrum., 76, 026104, https://doi.org/10.1063/1.1841972, 2005. 

Plainchont, B., Berruyer, P., Dumez, J. N., Jannin, S., and Giraudeau, P.: Dynamic Nuclear Polarization Opens New Perspectives for NMR Spectroscopy in Analytical Chemistry, Anal. Chem., 90, 3639–3650, https://doi.org/10.1021/acs.analchem.7b05236, 2018. 

Poole, C. P.: Electron spin resonance; a comprehensive treatise on experimental techniques, Interscience Publishers, New York, USA, xxvii, 922 pp., 1967. 

Rankin, A. G. M., Trébosc, J., Pourpoint, F., Amoureux, J.-P., and Lafon, O.: Recent developments in MAS DNP-NMR of materials, Solid State Nucl. Magn. Reson., 101, 116–143, https://doi.org/10.1016/j.ssnmr.2019.05.009, 2019. 

Reese, M., Lennartz, D., Marquardsen, T., Höfer, P., Tavernier, A., Carl, P., Schippmann, T., Bennati, M., Carlomagno, T., Engelke, F., and Griesinger, C.: Construction of a Liquid-State NMR DNP Shuttle Spectrometer: First Experimental Results and Evaluation of Optimal Performance Characteristics, Appl. Magn. Reson., 34, 301–311, 2008. 

Reese, M., TuÃàrke, M.-T., Tkach, I., Parigi, G., Luchinat, C., Marquardsen, T., Tavernier, A., HoÃàfer, P., Engelke, F., Griesinger, C., and Bennati, M.: 1H and 13C Dynamic Nuclear Polarization in Aqueous Solution with a Two-Field (0.35 T/14 T) Shuttle DNP Spectrometer, J. Am. Chem. Soc., 131, 15086–15087, https://doi.org/10.1021/ja905959n, 2009. 

Rosay, M., Blank, M., and Engelke, F.: Instrumentation for solid-state dynamic nuclear polarization with magic angle spinning NMR, J. Magn. Reson., 264, 88–98, https://doi.org/10.1016/j.jmr.2015.12.026, 2016. 

Sezer, D.: Rationalizing Overhauser DNP of nitroxide radicals in water through MD simulations, Phys. Chem. Chem. Phys., 16, 1022–32, https://doi.org/10.1039/c3cp53565g, 2014. 

Sharma, M., Janssen, G., Leggett, J., Kentgens, A. P., and van Bentum, P. J.: Rapid-melt Dynamic Nuclear Polarization, J. Magn. Reson., 258, 40–48, https://doi.org/10.1016/j.jmr.2015.06.007, 2015. 

Song, Y.-Q. and Kausik, R.: NMR application in unconventional shale reservoirs – A new porous media research frontier, Prog. Nucl. Magn. Reson. Spectrosc., 112–113, 17–33, https://doi.org/10.1016/j.pnmrs.2019.03.002, 2019. 

Song, Y.-Q., Souza, A., Vembusubramanian, M., Tang, Y., Fellah, K., Feng, L., and Reeder, S. L.: Multiphysics NMR correlation spectroscopy, J. Magn. Reson., 322, 106887, https://doi.org/10.1016/j.jmr.2020.106887, 2021. 

Thrippleton, M. J., Edden, R. A. E., and Keeler, J.: Suppression of strong coupling artefacts in J-spectra, J. Magn. Reson., 174, 97–109, https://doi.org/10.1016/j.jmr.2005.01.012, 2005. 

Traficante, D. D. and Nemeth, G. A.: A new and improved apodization function for resolution enhancement in NMR spectroscopy, J. Magn. Reson., 71, 237–245, https://doi.org/10.1016/0022-2364(87)90053-9, 1987. 

Traficante, D. D. and Rajabzadeh, M.: Optimum window function for sensitivity enhancement of NMR signals, Concepts Magn. Reson., 12, 83–101, https://doi.org/10.1002/(SICI)1099-0534(2000)12:2<83::AID-CMR3>3.0.CO;2-H, 2000. 

Überrück, T., Neudert, O., Kreuer, K.-D., Blümich, B., Granwehr, J., Stapf, S., and Han, S.: Effect of nitroxide spin probes on the transport properties of Nafion membranes, Phys. Chem. Chem. Phys., 20, 26660–26674, https://doi.org/10.1039/C8CP04607G, 2018.  

Vu, T. N. and Laukens, K.: Getting Your Peaks in Line: A Review of Alignment Methods for NMR Spectral Data, Metabolites, 3, 259–276, https://doi.org/10.3390/metabo3020259, 2013. 

Webb, A.: Small-Volume Hyphenated NMR Techniques, in: Micro and Nano Scale NMR, Wiley-VCH Verlag GmbH & Co. KGaA, Weinheim, Germany, 353–379, https://doi.org/10.1002/9783527697281.ch12, 2018. 

Windt, C. W., Soltner, H., van Dusschoten, D., and Blümler, P.: A portable Halbach magnet that can be opened and closed without force: The NMR-CUFF, J. Magn. Reson., 208, 27–33, https://doi.org/10.1016/j.jmr.2010.09.020, 2011. 

Zangger, K.: Pure shift NMR, Prog. Nucl. Magn. Reson. Spectrosc., 86–87, 1–20, https://doi.org/10.1016/j.pnmrs.2015.02.002, 2015. 

Download
Short summary
Typically, low-field Overhauser dynamic nuclear polarization (ODNP) experiments are 1D NMR experiments to study hydration dynamics. Here, we demonstrate the application of ODNP-enhanced 2D J-resolved (JRES) spectroscopy to improve spectral resolution beyond the limit imposed by the line broadening introduced by the paramagnetic polarizing agent. Crucial to these experiments is interleaved spectral referencing to compensate for temperature-induced field drifts over the course of the experiment.