Workflow for systematic design of electrochemical in operando NMR cells by matching B0 and B1 field simulations with experiments

Michael Schatz, Matthias Streun, Sven Jovanovic, Rüdiger-A. Eichel, and Josef Granwehr
Abstract

Combining electrochemistry (EC) and nuclear magnetic resonance (NMR) techniques has evolved from a challenging concept to an adaptable and versatile method for battery and electrolysis research. Continuous advancements in NMR hardware have fostered improved homogeneity of the static magnetic field, B0, and the radio frequency field, B1, yet fundamental challenges caused by introducing essential conductive components into the NMR sensitive volume remain. Cell designs in EC–NMR have largely been improved empirically, at times supported by magnetic field simulations. To propel systematic improvements of cell concepts, a workflow for a qualitative and semi-quantitative description of both B0 and B1 distortions is provided in this study. Three-dimensional finite element method (FEM) simulations of both B0 and B1 fields were employed to investigate cell structures with electrodes oriented perpendicular to B0, which allow realistic EC–NMR measurements for battery and electrolysis applications. Particular attention is paid to field distributions in the immediate vicinity of electrodes, which is of prime interest for electrochemical processes. Using a cell with a small void outside the electrochemical active region, the relevance of design details and bubble formation is demonstrated. Moreover, B1 amplifications in coin cells provide an explanation for unexpectedly high sensitivity in previous EC–NMR studies, implying the potential for selective excitation of spins close to electrode surfaces. The correlation of this amplification effect with coin geometry is described by empirical expressions. The simulations were validated experimentally utilising frequency-encoded 1H profile imaging and chemical shift imaging of 1H, 13C, and 23Na resonances of NaHCO3 electrolyte. Finally, the theoretical and experimental results are distilled into design guidelines for EC–NMR cells.

1 Introduction

As a versatile tool for non-invasive and quantitative identification of molecules and their chemical environment, nuclear magnetic resonance (NMR) spectroscopy presents an advantageous partner to electrochemistry (EC) techniques. The synergy between these methods enables a comprehensive study of processes and reactions within an operating electrochemical cell, especially when EC methods struggle with ambiguity in the presence of side reactions (Lozeman et al.2020). Particularly, the integration of magnetic resonance imaging (MRI) and relaxometry has yielded valuable insights into the chemistry inside operational electrochemical cells (Krachkovskiy et al.2020; Jovanovic et al.2023). Unlike optical methods that often focus on electrode surfaces, NMR spectroscopy enables simultaneous investigations of both the electrode surface and the electrolyte composition (Borzutzki and Brunklaus2017).

However, in comparison to other spectroelectrochemical methodologies, the potential of EC–NMR techniques has not been fully exploited. Integrating electrodes, wires, and electronic components within the NMR sensitive volume is challenging, as this leads to distortions in the static magnetic field B0 and the radio frequency (RF) field B1, critical for precise NMR measurements (Pietrzak et al.2021). Therefore, in operando NMR is often a tug of war between the intrinsic conductive components of an electrochemical cell and the desire for high field homogeneity, which affect NMR signal sensitivity, resolution, and the quantitativeness of the method.

Numerous EC–NMR studies have tackled these challenges in the field of battery research due to the involvement of NMR-active nuclei in redox reactions crucial for battery electrochemistry (Krachkovskiy et al.2020; Pecher et al.2017; Hu et al.2016). A wide variety of cell concepts have been tailored to fit the basic requirements of EC–NMR (Chandrashekar et al.2012; Bhattacharyya et al.2010; Krachkovskiy et al.2016; Kayser et al.2018). Tang et al. (2021) provide an extensive review of MRI techniques applied to such battery cells. However, the nature and magnitude of B0 and B1 field distortions have remained a topic that is less studied.

To minimise the effects of these distortions, significant effort has been directed towards well-conceived cell designs. Basic requirements are that metallic parts should be non-magnetic and should not encase the cell. Otherwise, RF irradiation is shielded from the sample depending on the thickness and conductivity of the metal, described by the skin effect (Britton2014). At the same time, electrical contact has to be ensured, the cell has to be completely sealed to prevent leakage, and the filling factor should be sufficient (Tang et al.2021). Shielding due to the skin effect can be minimised by avoiding RF penetration through conductive components in order to excite the sample. Still, there are two main types of distortions that cannot be entirely inhibited. First, magnetic susceptibility gradients between metals and their surrounding materials result in inhomogeneous B0. Secondly, eddy currents are induced in metals by RF fields, leading to a spatially and temporally dependent alteration of the B1 field strength. This modulation leads to a loss in quantitativeness and altering of the circuit tuning (Britton2014; Vashaee et al.2015; Mohammadi and Jerschow2019).

Despite the hurdles posed by these B0 and B1 field distortions, several EC–NMR cell designs have been presented over the last decades (Mohammadi and Jerschow2019; Pietrzak et al.2021). The pioneering work of Richards and Evans (1975), where wires just outside the sensitive volume in a standard 5 mm NMR tube were deployed as electrodes, was the basis for studies several decades later, such as the work of Silva et al. (2019). They omitted sample rotation, as NMR hardware has improved in terms of B0 homogeneity, and mixing by external pumping was replaced by making use of the magnetohydrodynamic effect at high magnetic field strength. This effect arises from the Lorentz force acting on ions in solution, resulting in a convective flow that resembles an internal mixer in an in operando NMR cell (Benders et al.2020).

Several other types of cells were developed that aimed at mitigating field distortions by minimising metal content in the cell, by either thin metallic films or foils (Webster2004; Zhang and Zwanziger2011; Jovanovic et al.2021; Schatz et al.2023), electrodes placed outside the sensitive volume (Hallberg et al.2008), or non-metallic electrodes of low conductivity (Klod et al.2009; Bussy et al.2013). Despite these advancements, the quantitative prediction of B0 and B1 field distortions remained challenging. Understanding the nature and magnitude of these distortions is critical, as it facilitates quantitative data evaluation or tailoring of pulse sequences, e.g. via techniques such as optimal control.

In the few selective studies found in the literature on the topic of understanding and mitigating field distortions in EC–NMR, the main focus is typically the skin effect and B1 distortions. Britton (2014) examined the influence of electrode orientation on the B1 field, employing 1H MRI to study the electrolyte surrounding a metal strip. Minimised distortions were achieved by parallel alignment of the metal relative to the orientation of RF irradiation (Mohammadi and Jerschow2019). Jovanovic et al. (2021) confirmed this using nutation experiments with varying angle between the electrode and the B1 field. To comprehend the skin effect in a Li electrode at varying B1 field directions, Ilott et al. (2014) conducted comprehensive calculations and validated their findings utilising 7Li MRI. Their investigations expanded into the influence of B0 field variations on the chemical shift of Li metal, arising from susceptibility effects and the Knight shift. Vashaee et al. (2015) employed SPRITE pulse sequences that are largely immune to susceptibility and gradient effects. By linearly increasing the flip angles, the signal intensity increases with B12, which is utilised to determine the spatial distribution of B1 precisely (Vashaee et al.2013). This allowed for distinct investigation of the eddy current effect of RF excitation near metal strips, emulating electrodes used in electrochemical applications. Their experimental findings were validated by B1 field simulations in electrode proximity. Numerous combinations of relative B0 and B1 field orientation, imaging pulse sequences, and cell geometries using flat electrodes were tested by Serša and Mikac (2018) and compared in terms of signal intensity. Eddy currents were identified as the main reason for signal loss. Thus, a parallel orientation of B1 to the electrode was found to be optimal, while the influence of B0 field orientation was minor. Purely phase-encoded single-point imaging showed the best performance.

Zhang et al. (2011) employed this to their advantage by implementing a parallel-plate resonator in an imaging probe for the investigation of flat membranes in a fuel cell setup. By maximising the filling factor and minimising B1 field distortions, high-resolution imaging could be applied to thin films despite the presence of auxiliary conductive components. This was achieved by not only considering a minimum of RF shielding and the optimal position of the B1 field parallel to the electrode surface but also inhibiting magnetic field gradient direction impinging perpendicularly to conductive surfaces. This cell concept minimises distortions caused by all interaction of conductive materials with electromagnetic fields, i.e. the B0 field, B0 field gradients, and the B1 field. It was applied and investigated both in simulations and in experiments for lithium ion battery studies (Aguilera et al.2021; Sanders et al.2022). Simulations by Walder et al. (2021) showed that between electrodes placed parallelly to B1, no significant decrease in RF field was expected, even inside a metal casing typical for industrial coin cells. Subsequently, they demonstrated first 7Li and 19F in operando NMR measurements of a commercial coin cell considering the calculated modulations of B0 and the B1 field.

While these studies have significantly contributed to comprehending RF field variations, there persists a need for workflows that include B0 and B1 field simulations of realistic cell setups alongside experimental validation. With the predominant focus on simulations and experimental determination of the B1 field, the significance of B0 fields often goes overlooked. Moreover, the utilisation of higher magnetic fields, beyond 2.4 T, which have been applied in previous studies, remains largely unexplored in this context.

In this study, a workflow that simultaneously considers B0 and B1 field distortions within an electrochemical NMR cell is presented. This approach involves simulating these fields using finite element methods (FEMs) and subsequently validating the findings experimentally through imaging techniques. Adapting the optimised EC–NMR cell for CO2 electroreduction from our previous studies (Schatz et al.2023), eddy currents were reduced. Utilising the B0 gradients provided by a diffusion probe, imaging along the axis perpendicular to the electrode was applied to determine local variations of B0 and B1 field along this specific dimension. Due to the symmetry of the cell setup, this was the most relevant dimension of the setup to probe spatial variations of 1H, 13C, or 23Na resonances. This not only allowed us to test whether B1 field distortions can really be considered minimal, but also even revealed signal enhancing effects. Regarding B0, field mapping using chemical shift imaging (CSI) of various nuclei is introduced. On the other hand, using this specific probe with the described symmetrical arrangement did not allow for further eddy current reduction in terms of B0 field gradient direction, as described by Zhang et al. (2011).

2 Experimental setup

A reference sample was prepared following the design of the electrochemical cell presented in our previous publication (Schatz et al.2023) that was aligned with the optimal cell designs outlined in the existing literature: the Cu working electrode (WE) was thin and oriented in parallel with B1 and perpendicularly to the B0 field and B0 field gradients. The reference electrode, counter electrode, the contacting wire, and the cell holder of the described setup were omitted. 13C-enriched 1 M NaHCO3 in aqueous solution was used as an electrolyte to enable 1H, 13C, and 23Na MRI. Figure 1a–b illustrate this sample, depicting two scenarios: one with the notch below the electrode filled with electrolyte and the other with air. The purpose of this notch in the electrolysis cell setup was facile contacting of the electrode using a Cu wire and optional mounting of the electrode inside it. The two scenarios were selected for investigation as electrolysis experiments were conducted in operando with air underneath the electrode. The air-filled notch could moreover be considered a model for gas entrapment in the cell housing or gas bubbles forming during electrolysis.

https://mr.copernicus.org/articles/5/167/2024/mr-5-167-2024-f01

Figure 1(a–b) Sectional side view of experimental setup for investigations of changes in the B0 field without and with electrolyte in the notch underneath the Cu WE, respectively. (c) Technical drawing with exact dimensions given.

Download

A Bruker DiffBB BBO broadband diffusion probe, which offers magnetic field gradients along the z axis (direction of B0, as shown in Fig. 1a), was operated on a Bruker Avance III HD spectrometer (Bruker BioSpin GmbH, Rheinstetten, Germany) with a 14.1 T wide-bore magnet. To distinguish between effects of cell and holder geometry on the B0 field and the influence of the electrode, shimming on the cell was performed without the Cu foil thoroughly. Instead of aiming for the minimum linewidth, the spatial homogeneity of B0 over the z dimension was prioritised during shimming and validated afterwards by 23Na CSI. 23Na linewidths of ca. 8 Hz and 1H linewidths of ca. 6 Hz could be achieved. The Cu electrode was subsequently added to the cell, either with air or with electrolyte in the cylindrical notch underneath. Without further adjustments of the magnetic field, purely phase-encoded CSI was performed, as described previously (Schatz et al.2023). The field of view was set to 20 mm. With 128 points in the spatial dimension and the zero filling to 256 points, the spatial resolution was 78 µm. Here, the 1H resonance of water and the 13C and 23Na resonance of the NaHCO3 electrolyte were imaged along the z axis. Comparable local changes in chemical shift across all three nuclei would indicate a nucleus-independent chemical shift (NICS), implying a purely magnetic field effect. Additionally, a frequency-encoded 1H profiling pulse sequence, also described previously (Schatz et al.2023), was performed. With 664 points in a field of view of 20 mm, the spatial resolution was 30 µm. Here, a flip angle below 90° was employed, such that a locally increased B1 field would result in increased intensity.

To validate the observations, FEM simulations of the setup shown in Fig. 1 were conducted using the magnetostatic simulator of the software FEMM (Version 4.2, David Meeker). For the electrode, material properties of copper were chosen, and for the surrounding liquid pure water was assumed. The magnetic susceptibility of PEEK was set to -9.335×10-6 (Wapler et al.2014). Before the simulation the boundary conditions of a volume with cylindrical geometry were optimised in order to obtain a uniform magnetic field. This corresponds to shimming of the magnet in the experiment and could be done for an empty volume but also for a volume that is already populated with parts of the setup, e.g. the PEEK holder and liquid without the electrode. Next, simulations were performed for the setup with and without the Cu electrode. The change in B0 was obtained as the ratio of the calculated field distributions.

To investigate the B1 field variations in a coin cell for battery applications, a sample imitating such a cell was prepared as shown in Fig. 2a–c. The thickness of the Cu discs was 1 or 5 mm, and the distance between the discs was 0.1 or 1 mm. All four possible combinations of disc thickness and distance were tested. To fix the distance, ring spacers made of PEEK were utilised; see Fig. 2. The volume between the discs was filled with high-performance liquid chromatography water (Sigma Aldrich Chemie GmbH, Taufkirchen, Germany). Here, no 13C and 23Na measurements were required. Thus, water was used instead of an electrolyte containing 13C and 23Na species. To exclude water from the sensitive volume above and below the coin cell, solid PEEK cylinders were used to fill this space.

https://mr.copernicus.org/articles/5/167/2024/mr-5-167-2024-f02

Figure 2(a–b) Sectional view of the experimental coin cell setup for investigations of changes in the B1 field with PEEK and Cu discs, respectively. (c) Technical drawing of the sectional view with exact dimensions given. (d) Setup in FEM simulations using EMpro showing the RF coils and the resulting B1 field direction parallel to the discs.

Download

Variation of the B1 field between the coins was determined by nutation experiments. The evolution of the water resonance intensity was investigated, while the pulse length was linearly increased at constant pulse power. First, the nutation frequency was determined for a sample without conductive metal discs. Instead, PEEK coins of the same dimension were introduced. This is subsequently compared to the nutation frequency with Cu coins in the sample.

The distortion of the B1 field surrounding the metal electrodes was calculated by numerical simulation using EMpro (Version 2020, Keysight Technologies). To mimic a uniform RF field close to the electrodes, a square Helmholtz coil was included in the simulation, comprising two parallel square-shaped wires spaced at 0.5445 times the length of each side of the square (Heller1955). The coil material was assumed to be an ideal conductor, and both square sections of the coil were concurrently driven by a current source. Two discs representing the coin cell were placed centrally between the Helmholtz coils. The dimensions of the discs and the distance between them were varied according to the variations in the experiment, while the material properties of the coins were those of either PEEK or Cu. As the wavelength of RF fields in high-field NMR is on the order of metres, while geometries in EC–NMR cells are on the orders of millimetres, it was assumed that magnetostatic laws can be applied; thus simulations were performed time-independently (Walder et al.2021).

Analogously to the FEM simulations described above using EMpro, a copper disc, with identical dimensions to the Cu electrode in Fig. 1 (4 mm diameter, 50 µm thickness), was positioned centrally between the two Helmholtz coils. A local increase in B1 field intensity results in an increased flip angle (Hoult and Richards1976). Thus, such a locally increased B1 field was detectable as a peak in the frequency-encoded 1H profiles if excitation was conducted using a flip angle of less than 90° for the bulk electrolyte.

3 Results and discussion

Compared to an NMR tube filled only with liquid electrolyte, the insertion of the PEEK holder already introduces a nucleus-independent degradation of NMR spectra due to susceptibility differences between the liquid and polymer, causing B0 distortions. Therefore, the B0 field was homogenised by shimming after insertion of the PEEK holder but before introduction of any conductive material. As a result, a high spatial homogeneity of the NMR spectra with maximum 0.05 ppm deviation of the peak position and linewidths of 8 Hz for 23Na could be achieved. The corresponding 1H, 13C, and 23Na CSI spectra after shimming are depicted for z=[0,2] mm in Fig. 3a–c, respectively, which are used as a reference for all further experiments.

Afterwards, the Cu foil was inserted while maintaining the position of the cell relative to the centre of the sensitive NMR volume. As a result, local changes in the chemical shifts for all three nuclei were observed, which are depicted in Fig. 3e–g for the case of air and Fig. 3i–k for the case of electrolyte in the notch underneath the Cu foil. In general, peak splitting and shifting of the same magnitude on the scale of parts per million (ppm) were observed for 1H, 13C, and 23Na measurements, indicating that this effect is nucleus-independent and can be attributed to magnetic field effects. Interestingly, significant differences in the extent of magnetic field alterations were observed between the two scenarios of air and electrolyte underneath the Cu WE. In the case of an air-filled notch, the resonance developed both a new, ca. 3 ppm downfield-shifted feature and a ca. 1 ppm upfield-shifted feature in electrode proximity. With increasing distance to the electrode, the two features converged, resulting in the bulk resonance that is significantly broadened, even at 2 mm distance to the Cu foil. In the case of an electrolyte-filled notch, no peak splitting was observed. Instead, the single peak was shifted to lower frequencies by ca. 0.5 ppm. In both scenarios and for all nuclei, overall decreased signal intensities were observed in spectra corresponding to volumes close to the Cu foil, especially for z=[0,0.5] mm. This is expected in CSI measurements near paramagnetic electrons on electrode surfaces with additional distortions by gradient pulses and echo formation (Bazak et al.2020). In quantitative measurements, this disruption of CSI images can be corrected by calibration of the intensity profile using pulse sequences with small angle excitation and stabilised gradients (Sethurajan et al.2015; Krachkovskiy et al.2016; Serša and Mikac2018; Bazak et al.2020). Lastly, the positioning of the Cu foil with respect to the z axis and the insertion of the tube were not perfectly reproducible and, therefore, have to be considered potential error sources.

https://mr.copernicus.org/articles/5/167/2024/mr-5-167-2024-f03

Figure 3Spatially resolved NMR spectra of CSI measurements of experimental setups (a–c) without Cu foil, (e–g) with Cu foil and air underneath, and (i–k) with Cu foil and electrolyte underneath, depicted between z=[0,2] mm. Spectra are shown for 1H, 13C, and 23Na, respectively. Simulated B0 distortions, represented by histograms over the same z interval, for the case of (d) no Cu foil and water filled notch, (h) air, and (l) water underneath the Cu foil, respectively. The histograms were corrected under the assumption of perfect shimming of the sample without Cu foil.

Download

To compare the simulation results to these spatially resolved NMR spectra, distributions of B0 variation from excitation frequency, depicted as histograms, were calculated for all voxels in slices of 100 µm thickness in the z direction. The simulations for z=[0,2] mm are shown in Fig. 3h and l with air and water underneath the electrode, respectively. A reference calculation without Cu foil and with a water filled notch was used to correct these histograms, assuming a perfect shim on the sample without conductive material. The weighted average of each histogram of this simulation was subtracted from the respective histogram representing the same slices. The results of all three simulation scenarios were corrected accordingly and are depicted in Fig. 3d. Uncorrected histograms are given in Fig. A1 of Appendix A. The simulated frequency distributions are in good agreement with the experimental CSI spectra. For the air-filled notch, a splitting of ca. 4 ppm for the two spectral features was observed, which decreases to zero within 2 mm along the z axis. The upfield peak predicted by simulations was shifted by ca. 3 ppm, while the downfield peak was shifted by ca. 1 ppm, which is in quantitative accordance with experiments. For the case of water under the electrode, the simulation showed no significant difference to the reference simulation without Cu. Thus, the simulation results reflected experimental findings both qualitatively and quantitatively; however a post-processing correction had to be applied assuming a perfect shim. Even though uncorrected, the B0 distortions, depicted in Fig. A1, show good qualitative agreement with experimental data. Here, the calculated values of the distorted signals were shifted upfield compared to experimental results, which could be attributed to an unshimmed magnetic field. Shimming could only be carried out by post-simulation data processing, and the minimised discrepancy between experimental results and corrected simulations validates this processing step.

Overall, the agreement of simulations and experiments demonstrates that CSI is capable of probing the influence of conductive cell components on B0 and that FEM calculations are suitable to predict the spatially resolved magnetic field distortions in the proximity of conductive materials and their effect on NMR spectra.

The results from combined experiments and simulations indicated that a conductive electrode in the NMR sensitive volume will induce a systematic error in imaging and spectroscopy experiments. By careful shimming after inserting the full cell setup including all electrodes and wires, this error may be reduced. However, it can not be ruled out that a chemical shift alteration near the electrode remains. Before every in operando measurement, the change in chemical shift that remains after shimming has to be determined and, subsequently, corrected. In our previous study, the peak splitting illustrated in Fig. 3e–g could be homogenised to a single peak with linewidth on the order of ca. 6 Hz by shimming. A chemical shift difference between the near electrode and bulk electrolyte remained even after a long equilibration time and thorough shimming and had to be corrected during post-processing (Schatz et al.2023). In particular, local differences to the bulk chemical shift in the reference CSI measurements were subtracted from chemical shift profiles acquired during electrolysis.

The greater impact of the air pocket compared with an electrolyte-filled notch not only highlights the necessity to minimise entrapped air in EC–NMR cell setups and reduce the number of materials with high susceptibility differences but also mitigates the evolution of gas bubbles during electrolysis. While the inhomogeneity caused by air pockets can mostly be corrected by shimming before the electrolysis, the gas bubble distortions can not be shimmed during electrolysis as they build up, grow, and potentially detach in irregular patterns. If gas bubble formation cannot be inhibited, a robust shim is required, and acquisition time has to be sufficiently long to average out errors over time. Moreover, multi-nuclear measurements can be utilised to correct for the nucleus-independent chemical shift effects of gas bubbles during post-processing.

The total intensity of 1H density from frequency-encoded images is plotted over the z axis in Fig. 4d for the experimental setup with an air-filled notch. Here, a pulse length of 16.5 µs was used, while the pulse duration for a 90° magnetisation flip was determined to be 18.5 µs. Thus, the increased intensity close to the electrode at z=0 mm is attributed to a locally enhanced B1 field. Towards positive z values, the 1H density approaches zero at z=15 mm, delineating the upper boundary of the sensitive volume. In the negative z range, low 1H density indicates the position of the PEEK holder that does not yield a comparably high 1H signal intensity as the liquid electrolyte. The residual 1H signal and the peak at ca. z=−1 mm result from thin films of electrolyte on the surface of the PEEK holder.

Figure 4a–c present the outcome of three-dimensional B1 field simulations utilising EMpro. The magnitude of B1 field was calculated considering all components of the three-dimensional vector. The resulting B1 field intensities are illustrated in the top view of the Cu disc, exactly on top of the disc and 50 µm above its surface, alongside a sectional view through the disc's centre. Notably, the simulations confirm a localised increase in the B1 field near the WE by about 20 % compared to the bulk induced by eddy currents, primarily evident at the edges of the Cu foil. Concurrently, B1 field distortions extend less than 50 µm into the z direction, predominantly confined to volumes close to the electrode edges. Depending on the exact cell geometry and setup, this effect requires consideration in quantitative experiment evaluations. Inhomogeneities in the static B0 field could also be accounted for alterations in the frequency-encoded profile. As the sample was thoroughly shimmed, this contribution to the image distortions was neglected. However, for the cell setup utilised in this work, the associated error from B1 field distortions is estimated to be relatively minor. In recent imaging experiments with purely phase-encoded pulse sequences, the spatial resolution was on the order of 100 µm and could therefore not resolve this effect. In this case, the field of view of the image would be altered instead of the cell geometry itself.

https://mr.copernicus.org/articles/5/167/2024/mr-5-167-2024-f04

Figure 4(a–b) B1 field intensities in layers exactly on top of the Cu disc and 50 µm above it, respectively. (c) B1 field intensities in the sectional side view through the Cu disc. The magnitude of B1 field was calculated considering all components of the three-dimensional vector. (d) Frequency-encoded 1H density profile of the experimental setup with horizontally placed Cu foil. For the imaging experiment, a pulse length of 16.5 µs was used, while the 90° pulse duration was 18.5 µs. Thus, increased intensities close to the electrode at z=0 mm can be correlated to local enhancement of B1 field.

Download

In the case of a second conductive disc added to the B1 field simulations to mimic a coin cell, a sectional view through the centre of the discs is depicted in Fig. 5 for 1 and 0.1 mm distance and 5 and 1 mm coin thickness, respectively. Notably, a uniform increase in B1 field intensity was observed between the coins in all cell configurations. With increasing coin thickness and decreasing distance between coins, this effect is further amplified. The B1 field enhancement is evident in the histogram in Fig. 5e, where the four scenarios are compared to a reference case without any conductive material in the simulation volume. All four simulations showed a high homogeneity in distribution of enhanced B1 field. Further examination of the B1 field along the x and y direction parallel to the disc surface on a line precisely in the centre between the two discs, as depicted in Fig. 6a for the four aforementioned combinations of coin thickness and distance, quantified the two overarching trends: an increase in the B1 field with (1) decreasing distance and (2) increasing thickness of discs. The percentage increase in B1 intensity between the electrodes compared to bulk values, as determined from simulations, is contrasted to the experimentally measured increase in nutation frequency in Table 1. The experimental value was determined by assessing the nutation frequency of the water resonance between non-conductive PEEK discs and conductive Cu discs; see Fig. 2a and b, respectively. Details about the determination of nutation frequencies are given in Appendix B. Given that the nutation frequency is directly proportional to B1 field, the accordance between simulated and experimental data substantiates that coin cells in parallel orientation to the B1 field exhibit a significant B1 amplification for the volume in between. Compared with other studies calculating B1 intensity between parallel plates (Zhang et al.2011; Walder et al.2021), where no RF field intensification was observed, setups with larger electrode thickness and smaller distance were investigated, which may not correspond to dimensions of commercial coin cells but helped to uncover the amplifying effect and enabled its measurement by nutation experiments.

To extrapolate the extent of this effect, additional calculations were conducted with the coin cell arrangements that led to maximum B1 amplification, i.e. with a coin thickness of 5 mm and a minimum distance of 0.05 mm, where the distance is on the order of realistic coin cell dimensions. While the extreme value of either coin thickness or distance was kept constant, the other respective value was varied in the interval [0.05, 0.1, 0.5, 1, 2, 3, 4, 5] mm. The RF amplification is displayed as a function of the varied value for the two extreme cases in Fig. 6b. The B1 profile between the coins in parallel direction to the coin surface is displayed in Fig. B3 in Appendix B. The cell configuration with 0.05 mm thickness and distance that is best comparable to commercial coin cells exhibited an RF amplification of less than 5 %, which could easily be overlooked or interpreted to be within error margins when measuring nutation frequency in a coin cell for in operando NMR investigations (Kayser et al.2018; Walder et al.2021). Even though electrode thicknesses on the order of 1 mm are not realistic for commercial cells, this effect could be utilised for in operando NMR cells with amplification of B1 field of 90 % and more. With increasing coin thickness or distance, the amplification effect increases or decreases exponentially, respectively, evolving towards a limit. Exponential fits for the two simulation series are plotted as dashed lines in Fig. 6b, described by

(1) a ( δ ) = 74.2 % e - 0.472 mm - 1 δ , R 2 = 0.976 ,

with RF amplification a (in %) and coin distance δ at a constant thickness of 5 mm, and

(2) a ( d ) = 77.5 % ( 1 - e - 0.707 mm - 1 d ) , R 2 = 0.997 ,

with coin thickness d at a constant distance of 0.05 mm, respectively. From these two correlations, a unified equation could be found,

(3) a ( δ , d ) = 77.9 % e - 0.472 mm - 1 δ ( 1 - e - 0.707 mm - 1 d ) ,

which is derived in Appendix C.

Eddy currents that are induced in the conductive material perpendicularly to the B1 direction cause a circulating current along the coin surface. The currents on the surfaces of the two coins facing each other cause modulations of the B1 field, but in contrast to the scenario with just one coin, these surface currents are aligned in the same direction and therefore cause spatially uniform amplification of B1. Resonance effects between the coins comparable to a waveguide effect could potentially be accountable for the RF amplifying phenomenon, though further investigation via time-resolved simulations would be required to elucidate the underlying processes definitively. Regardless of the exact mechanism taking place, the closer the amplifying currents are located to each other, i.e. with decreasing coin distance, the greater the intensification of RF field. Increased thickness of the conductive discs has two effects, further increasing eddy currents along the surfaces facing each other. First, with increasing thickness the overall eddy current strength is enhanced proportionally to the interface of conductive material perpendicular to B1. Secondly, in a thicker coin, the opposing surface currents are further apart, which decreases attenuation between them.

An additional simulation of B1 field distribution of the coin cell setup is depicted in Fig. D1 of Appendix D, in which the coins were directly employed as part of the resonant circuit. Here, the excitation patterns were fundamentally different from excitation using an external coil, which provides an approach to explain the non-intuitive correlation of B1 amplification with coin thickness. The B1 distribution was strongly influenced by eddy currents at the edges of the coin, which created homogeneous B1 fields inside the coin cell structure. However, when excitation occurs directly through the coins, different eddy currents occur, and this homogeneity collapses, and instead, highly localised B1 amplifications occur. In contrast to external excitation, coin thickness variations did not lead to significant differences in B1 distributions.

https://mr.copernicus.org/articles/5/167/2024/mr-5-167-2024-f05

Figure 5Spatial distribution of calculated B1 field in a central sectional plane perpendicular to the coin surface with coins of (a) 1 mm distance and 1 mm thickness, (b) 0.1 mm distance and 1 mm thickness, (c) 1 mm distance and 5 mm thickness, and (d) 0.1 mm distance and 5 mm thickness. The origin of the z axis is centred between the two coins, whose edges are marked by white, dashed lines. (e) Histogram of B1 field intensities between the coins for the four presented calculations. A simulation without conductive coins (see “empty”) serves as a reference case to quantify B1 field enhancement. All histograms were normalised; therefore all integrals are equal.

Download

https://mr.copernicus.org/articles/5/167/2024/mr-5-167-2024-f06

Figure 6(a) Calculated B1 field from simulations in a line along the x axis, parallel to and central between the coin surfaces. The calculated B1 field in an empty volume without conductive coins is plotted as a reference. (b) RF amplification in comparison to B1 field in an empty volume for two extreme conditions: For a coin distance of 0.05 mm, thickness was varied between [0.05, 0.1, 0.5, 1, 2, 3, 4, 5] mm. For a coin thickness of 5 mm, distance was varied between [0.05, 0.1, 0.5, 1, 2, 3, 4, 5] mm. Dashed lines represent exponential fits to the data.

Download

Table 1Increase in the B1 field in simulations and experimentally determined increase in nutation frequency with conductive Cu discs compared to PEEK discs of 1 and 5 mm thickness and 1 and 0.1 mm distance, respectively.

Download Print Version | Download XLSX

4 Conclusions

This study compared predictive simulation methods and their experimental validation for assessing B0 and B1 field distortions induced by electrically conductive components within the NMR sensitive volume, with emphasis of EC–NMR application. FEM simulations highlighted the significant impact of materials with magnetic susceptibility differences on B0 homogeneity and demonstrated semi-quantitative agreement with experimental B0 field distortions measured by 1H, 13C, and 23Na CSI. It was observed that the presence of an air pocket beneath the electrode posed challenges due to a large difference in magnetic susceptibility compared to the surrounding. Although this inhomogeneity was shown to be largely correctable by careful shimming, improved susceptibility matching of cell components could facilitate the B0 field homogenisation. Not only air pockets in the cell setup but also gas bubbles emerging during electrolysis operation should therefore be minimised or discharged from the sensitive volume.

FEM simulations of B1 field effects were in accordance with local increases in frequency-encoded 1H profiles. Between electrodes of coin cells, a uniformly enhanced B1 field was predicted by simulations and experimentally confirmed by nutation experiments. The intensification of RF field intensity was found to increase with decreasing electrode distance and increasing electrode thickness. This finding contributes to the understanding of why EC–NMR battery cells in previous studies (Kayser et al.2018; Walder et al.2021) have exhibited performance beyond expectations and how coin cells could be employed to their advantage in EC–NMR applications. As the B1 amplification is comparable to waveguide effects, further time-resolved investigations are suggested to understand the nature of this phenomenon. In this context, it is also important to note that the sample itself becomes an integral part of the resonant circuit of the probe. Once resonance structures are integrated into the NMR sensitive volume, focusing solely on the skin effect is not sufficient and will not yield quantitative results.

Future studies with higher spatial resolution or surface sensitivity should consider a locally inhomogeneous nutation frequency. This effect could even be employed to an advantage by slice selection inspired by the DANTE sequence (Maffei et al.1991) or by numerical pulses optimised by optimal control. Such tailored pulse sequences may mitigate magnetic field distortions or even enable the selective excitation of certain volumes or species otherwise not possible. As volumes close to the electrode are affected by increased B1 field, this method could differentiate between signals of bulk and near-surface electrolyte. Additional distorting effects caused by gradient switching during diffusion or imaging experiment, including concomitant gradients in high field applications, should likewise be considered in that regard. Thus, further advancements of the simulation should incorporate the geometry of coils and their electrical wiring, as well as, in a next step, the pulse sequence, which would require combining field simulations and spin dynamics.

Ultimately, all the design guidelines for EC–NMR cells given in this study and in the literature can not comprise a universally valid recipe but should be consulted in their context. Depending on the specific requirements of the research question, design adjustments could be guided towards either high sensitivity, by employing amplification effects or increased filling factor, or high field homogeneity, by the susceptibility matching of materials and sophisticated placement of conductive components. By utilising the presented workflow of matching B0 and B1 field simulation with imaging experiments, the quality of outcomes of in operando EC–NMR experiments can be aligned with the research objectives.

Appendix A:B0 field simulations without shim correction
https://mr.copernicus.org/articles/5/167/2024/mr-5-167-2024-f07

Figure A1Simulated B0 distortions, represented by histograms over the z axis, for the case of (a) no Cu foil and electrolyte-filled notch, (b) air, and (c) electrolyte underneath the Cu foil, respectively. The histograms were not corrected in terms of shimming.

Download

Appendix B: Nutation experiments

Table B1Integral intervals and nutation frequency determined for nutation experiments with PEEK and Cu discs of 1 and 5 mm thickness and 1 and 0.1 mm distance, respectively.

Download Print Version | Download XLSX

The nutation experiments were conducted with increasing pulse length in 80 steps between 5 and 400 µs. The delay between two experiments was 5 s, and acquisition time was 0.68 s. Spectra were processed with manual phase correction individually for every experiment, line broadening of 5 Hz, and no baseline correction. Signal integrals were plotted over pulse length to generate nutation curves. Integral intervals were chosen to fit the resonance frequency of the water between the coins. To distinguish between these resonances and signals originating from the water films between PEEK cylinders and NMR tube glass or in between PEEK parts, 1H CSI was employed providing the integral intervals in Table B1. Water from outside the gap between the coins showed a significant different chemical shift; thus the signal of water between the coins was well distinguishable. Only in the case of Cu discs with 0.1 mm distance and 1 mm thickness could the signal not be perfectly resolved, resulting in high inhomogeneity in the nutation curve and a second peak in the nutation spectrum. Nutation spectra were calculated by fast Fourier transform of the nutation signal and plotted in magnitude mode over positive frequency values. Resulting nutation frequencies are displayed in Table B1. Nutation curves and spectra of experiments with PEEK and Cu discs of 1 and 5 mm thickness and 1 and 0.1 mm distance are illustrated in Figs. B1 and B2.

https://mr.copernicus.org/articles/5/167/2024/mr-5-167-2024-f08

Figure B1(a, c, e, g) Nutation curves and (b, d, f, h) nutation spectra of the 1H water resonance in nutation experiments with coin cell setups with 1 mm electrode distance, respectively. (a–b) PEEK discs with 1 mm thickness. (c–d) Cu discs with 1 mm thickness. (e–f) PEEK discs with 5 mm thickness. (g–h) Cu discs with 5 mm thickness. The point with highest intensity in nutation spectra is selected as the nutation frequency. Numerically exactly matching values are due to the limited number of 80 discrete points of the nutation curve.

Download

https://mr.copernicus.org/articles/5/167/2024/mr-5-167-2024-f09

Figure B2(a, c, e, g) Nutation curves and (b, d, f, h) nutation spectra of the 1H water resonance in nutation experiments with coin cell setups with 0.1 mm electrode distance, respectively. (a–b) PEEK discs with 1 mm thickness. (c–d) Cu discs with 1 mm thickness. (e–f) PEEK discs with 5 mm thickness. (g–h) Cu discs with 5 mm thickness. The point with highest intensity in nutation spectra is selected as the nutation frequency. Numerically exactly matching values are due to the limited number of 80 discrete points of the nutation curve.

Download

https://mr.copernicus.org/articles/5/167/2024/mr-5-167-2024-f10

Figure B3Calculated B1 field from simulations in a central line along the x axis, parallel to the coin surface. (a) For a coin distance of 0.05 mm, thickness was increased in the order [0.05, 0.1, 0.5, 1, 2, 3, 4, 5] mm, indicated with changing colour from green to blue. (b) For a coin thickness of 5 mm, distance was decreased in the order [5, 4, 3, 2, 1, 0.5, 0.1, 0.05] mm, indicated with changing colour from green to blue.

Download

Appendix C: Derivation of a universal empirical equation for B1 field amplification as a function of coin thickness and distance

It is assumed that there exists a universal equation that equals Eqs. (1) and (2) and is described by

(C1) a ( δ , d ) = K e k δ δ ( 1 - e k d d ) ,

with kδ=-0.472 mm−1 and kd=-0.707 mm−1 from Eqs. (1) and (2), respectively, and a constant K that has to be determined. Equation (C1) has to match Eq. (1) for coin thickness d=5 mm,

(C2) a ( δ , d = 5 mm ) = K e k δ δ ( 1 - e k d 5 mm ) = ! 74.2 % e k δ δ , K = 74.2 % 0.971 = 76.4 % .

The same holds for Eq. (2) and coin distance of δ=0.05 mm,

(C3) a ( δ = 0.05 mm , d ) = K e k δ 0.05 mm ( 1 - e k d d ) = ! 77.5 % ( 1 - e k d d ) , K = 77.5 % 0.977 = 79.4 % .

As the two calculated values for the constant K differ only slightly, the average of these is used to obtain Eq. (3),

(C4) a ( δ , d ) = 77.9 % e - 0.472 mm - 1 δ ( 1 - e - 0.707 mm - 1 d ) .
Appendix D: Simulation of B1 field using coin cell electrodes for direct RF excitation
https://mr.copernicus.org/articles/5/167/2024/mr-5-167-2024-f11

Figure D1Calculated B1 field from simulations using the electrodes of coin cells as the RF coil. The magnitude of B1 field in the xz plane between coins with 1 mm distance is depicted for electrode thickness of (a) 1 mm and (b) 0.1 mm. (c) Mean field averaged over the z axis plotted over the x axis for electrode thickness of 1 mm and 0.1 mm.

Download

Code and data availability

TopSpin raw data of the presented measurements, as well as the results of B0 and B1 field simulations, are available on the Jülich DATA repository at https://doi.org/10.26165/JUELICH-DATA/KJTAXZ (Schatz et al.2024). The simulation codes used in this work and all other data are available from the authors upon request.

Author contributions

Each author contributed to this work as follows: MiS prepared the samples and performed the experiments in consultation with MaS, SJ, and JG; B0 and B1 field simulations were performed by MaS; data analysis and interpretation were performed by MS in collaboration with MaS, SJ, RAE, and JG; and the paper was written by MiS in collaboration with MaS, SJ, RAE, and JG. All the authors have read and agreed to the paper.

Competing interests

The contact author has declared that none of the authors has any competing interests.

Disclaimer

Publisher's note: Copernicus Publications remains neutral with regard to jurisdictional claims made in the text, published maps, institutional affiliations, or any other geographical representation in this paper. While Copernicus Publications makes every effort to include appropriate place names, the final responsibility lies with the authors.

Acknowledgements

The authors thank Christian Hellenbrandt for support in sample preparation.

Financial support

This research has been supported by the German Research Foundation (DFG) under Germany's Excellence Strategy–Cluster of Excellence 2186 “The Fuel Science Center” (grant no. 390919832). Josef Granwehr received funding from the German Federal Ministry of Education and Research (BMBF) as part of the research cluster “AQua” within the project “InOPlaBat” (grant no. 13XP0352B).

The article processing charges for this open-access publication were covered by the Forschungszentrum Jülich.

Review statement

This paper was edited by Jan Gerrit Korvink and reviewed by Bruce Balcom and two anonymous referees.

References

Aguilera, A. R., MacMillan, B., Krachkovskiy, S., Sanders, K. J., Alkhayri, F., Adam Dyker, C., Goward, G. R., and Balcom, B. J.: A parallel-plate RF probe and battery cartridge for 7Li ion battery studies, J. Magn. Reson., 325, 106943, https://doi.org/10.1016/j.jmr.2021.106943, 2021. a

Bazak, J. D., Allen, J. P., Krachkovskiy, S. A., and Goward, G. R.: Mapping of Lithium-Ion Battery Electrolyte Transport Properties and Limiting Currents with In Situ MRI, J. Electrochem. Soc., 167, 140518, https://doi.org/10.1149/1945-7111/abc0c9, 2020. a, b

Benders, S., Gomes, B. F., Carmo, M., Colnago, L. A., and Blümich, B.: In-situ MRI velocimetry of the magnetohydrodynamic effect in electrochemical cells, J. Magn. Reson., 312, 106692, https://doi.org/10.1016/j.jmr.2020.106692, 2020. a

Bhattacharyya, R., Key, B., Chen, H., Best, A. S., Hollenkamp, A. F., and Grey, C. P.: In situ NMR observation of the formation of metallic lithium microstructures in lithium batteries, Nat. Mater., 9, 504–510, https://doi.org/10.1038/nmat2764, 2010. a

Borzutzki, K. and Brunklaus, G.: Chapter Three – Magnetic Resonance Imaging Studies of the Spatial Distribution of Charge Carriers, in: Annual Reports on NMR Spectroscopy, edited by: Webb, G. A., Vol. 91, 115–141, Academic Press, ISBN 0066-4103, https://doi.org/10.1016/bs.arnmr.2016.12.003, 2017. a

Britton, M. M.: Magnetic resonance imaging of electrochemical cells containing bulk metal, Chem. Phys. Chem., 15, 1731–1736, https://doi.org/10.1002/cphc.201400083, 2014. a, b, c

Bussy, U., Giraudeau, P., Silvestre, V., Jaunet-Lahary, T., Ferchaud-Roucher, V., Krempf, M., Akoka, S., Tea, I., and Boujtita, M.: In situ NMR spectroelectrochemistry for the structure elucidation of unstable intermediate metabolites, Anal. Bioanal. Chem., 405, 5817–5824, https://doi.org/10.1007/s00216-013-6977-z, 2013. a

Chandrashekar, S., Trease, N. M., Chang, H. J., Du, L.-S., Grey, C. P., and Jerschow, A.: 7Li MRI of Li batteries reveals location of microstructural lithium, Nat. Mater., 11, 311–315, https://doi.org/10.1038/nmat3246, 2012. a

Hallberg, F., Furó, I., Yushmanov, P. V., and Stilbs, P.: Sensitive and robust electrophoretic NMR: instrumentation and experiments, J. Magn. Reson., 192, 69–77, https://doi.org/10.1016/j.jmr.2008.02.001, 2008. a

Heller, C.: Über die Erzeugung großräumiger homogener Magnetfelder zum Studium des Verhaltens von Magnetkompassen und Kompensiermitteln auf verschiedenen magnetischen Breiten, Deutsche Hydrografische Zeitschrift, 8, 157–164, https://doi.org/10.1007/BF02019812, 1955. a

Hoult, D. and Richards, R.: The signal-to-noise ratio of the nuclear magnetic resonance experiment, J. Magn. Reson., 24, 71–85, https://doi.org/10.1016/0022-2364(76)90233-X, 1976. a

Hu, J. Z., Zhao, Z., Hu, M. Y., Feng, J., Deng, X., Chen, X., Xu, W., Liu, J., and Zhang, J.-G.: In situ 7Li and 133Cs nuclear magnetic resonance investigations on the role of Cs+ additive in lithium-metal deposition process, J. Power Sources, 304, 51–59, https://doi.org/10.1016/j.jpowsour.2015.10.067, 2016. a

Ilott, A. J., Chandrashekar, S., Klöckner, A., Chang, H. J., Trease, N. M., Grey, C. P., Greengard, L., and Jerschow, A.: Visualizing skin effects in conductors with MRI: 7Li MRI experiments and calculations, J. Magn. Reson., 245, 143–149, https://doi.org/10.1016/j.jmr.2014.06.013, 2014. a

Jovanovic, S., Schleker, P. P. M., Streun, M., Merz, S., Jakes, P., Schatz, M., Eichel, R.-A., and Granwehr, J.: An electrochemical cell for in operando 13C nuclear magnetic resonance investigations of carbon dioxide/carbonate processes in aqueous solution, Magn. Reson., 2, 265–280, https://doi.org/10.5194/mr-2-265-2021, 2021. a, b

Jovanovic, S., Jakes, P., Merz, S., Daniel, D. T., Eichel, R.-A., and Granwehr, J.: In operando NMR investigations of the aqueous electrolyte chemistry during electrolytic CO2 reduction, Commun. Chem., 6, 268, https://doi.org/10.1038/s42004-023-01065-3, 2023. a

Kayser, S. A., Mester, A., Mertens, A., Jakes, P., Eichel, R.-A., and Granwehr, J.: Long-run in operando NMR to investigate the evolution and degradation of battery cells, Phys. Chem. Chem. Phys., 20, 13765–13776, https://doi.org/10.1039/C8CP01067F, 2018. a, b, c

Klod, S., Ziegs, F., and Dunsch, L.: In situ NMR spectroelectrochemistry of higher sensitivity by large scale electrodes, Anal. Chem., 81, 10262–10267, https://doi.org/10.1021/ac901641m, 2009. a

Krachkovskiy, S., Trudeau, M. L., and Zaghib, K.: Application of Magnetic Resonance Techniques to the In Situ Characterization of Li-Ion Batteries: A Review, Materials, 13, 1694, https://doi.org/10.3390/ma13071694, 2020. a, b

Krachkovskiy, S. A., Bazak, J. D., Werhun, P., Balcom, B. J., Halalay, I. C., and Goward, G. R.: Visualization of Steady-State Ionic Concentration Profiles Formed in Electrolytes during Li-Ion Battery Operation and Determination of Mass-Transport Properties by in Situ Magnetic Resonance Imaging, J. Am. Chem. Soc., 138, 7992–7999, https://doi.org/10.1021/jacs.6b04226, 2016. a, b

Lozeman, J. J. A., Führer, P., Olthuis, W., and Odijk, M.: Spectroelectrochemistry, the future of visualizing electrode processes by hyphenating electrochemistry with spectroscopic techniques, Analyst, 145, 2482–2509, https://doi.org/10.1039/c9an02105a, 2020. a

Maffei, P., Elbayed, K., Brondeau, J., and Canet, D.: Slice selection in NMR imaging by use of the B1 gradient along the axial direction of a saddle-shaped coil, J. Magn. Reson., 95, 382–386, https://doi.org/10.1016/0022-2364(91)90227-K, 1991. a

Mohammadi, M. and Jerschow, A.: In situ and operando magnetic resonance imaging of electrochemical cells: A perspective, J. Magn. Reson., 308, 106600, https://doi.org/10.1016/j.jmr.2019.106600, 2019. a, b, c

Pecher, O., Carretero-González, J., Griffith, K. J., and Grey, C. P.: Materials' Methods: NMR in Battery Research, Chem. Mater., 29, 213–242, https://doi.org/10.1021/acs.chemmater.6b03183, 2017. a

Pietrzak, M., Jopa, S., Mames, A., Urbańczyk, M., Woźny, M., and Ratajczyk, T.: Recent Progress in Liquid State Electrochemistry Coupled with NMR Spectroscopy, Chem. Electro. Chem., 8, 4181–4198, https://doi.org/10.1002/celc.202100724, 2021. a, b

Richards, J. A. and Evans, D. H.: Flow cell for electrolysis within the probe of a nuclear magnetic resonance spectrometer, Anal. Chem., 47, 964–966, https://doi.org/10.1021/ac60356a016, 1975. a

Sanders, K. J., Aguilera, A. R., Keffer, J. R., Balcom, B. J., Halalay, I. C., and Goward, G. R.: Transient lithium metal plating on graphite: Operando 7Li nuclear magnetic resonance investigation of a battery cell using a novel RF probe, Carbon, 189, 377–385, https://doi.org/10.1016/j.carbon.2021.12.082, 2022. a

Schatz, M., Kochs, J. F., Jovanovic, S., Eichel, R.-A., and Granwehr, J.: Interplay of Local pH and Cation Hydrolysis during Electrochemical CO2 Reduction Visualized by In Operando Chemical Shift-Resolved Magnetic Resonance Imaging, J. Phys. Chem. C, 127, 18986–18996, https://doi.org/10.1021/acs.jpcc.3c03563, 2023. a, b, c, d, e, f

Schatz, M., Streun, M., Jovanovic, S., Eichel, R.-A., and Granwehr, J.: Replication Data for: Workflow for Systematic Design of Electrochemical In Operando NMR Cells by Matching B0 and B1 Field Simulations with Experiments, V1, Jülich DATA [data set], https://doi.org/10.26165/JUELICH-DATA/KJTAXZ, 2024. a

Serša, I. and Mikac, U.: A study of MR signal reception from a model for a battery cell, J. Magn. Reson., 294, 7–15, https://doi.org/10.1016/j.jmr.2018.06.013, 2018. a, b

Sethurajan, A. K., Krachkovskiy, S. A., Halalay, I. C., Goward, G. R., and Protas, B.: Accurate Characterization of Ion Transport Properties in Binary Symmetric Electrolytes Using In Situ NMR Imaging and Inverse Modeling, J. Phys. Chem. B, 119, 12238–12248, https://doi.org/10.1021/acs.jpcb.5b04300, 2015.  a

Silva, P. F. d., Gomes, B. F., Lobo, C. M. S., Queiroz Júnior, L. H. K., Danieli, E., Carmo, M., Blümich, B., and Colnago, L. A.: Electrochemical NMR spectroscopy: Electrode construction and magnetic sample stirring, Microchem. J., 146, 658–663, https://doi.org/10.1016/j.microc.2019.01.010, 2019. a

Tang, M., Fu, R., Deschamps, M., Romanenko, K., and Jerschow, A.: The Methodology of Electrochemical In Situ NMR and MRI, in: NMR and MRI of Electrochemical Energy Storage Materials and Devices, edited by: Yang, Y., Fu, R., and Huo, H., 71–105, The Royal Society of Chemistry, ISBN 978-1-78801-848-7, https://doi.org/10.1039/9781839160097-00071, 2021. a, b

Vashaee, S., Newling, B., MacMillan, B., and Balcom, B. J.: B(1) mapping with a pure phase encode approach: quantitative density profiling, J. Magn. Reson., 232, 68–75, https://doi.org/10.1016/j.jmr.2013.04.012, 2013. a

Vashaee, S., Goora, F., Britton, M. M., Newling, B., and Balcom, B. J.: Mapping B(1)-induced eddy current effects near metallic structures in MR images: a comparison of simulation and experiment, J. Magn. Reson., 250, 17–24, https://doi.org/10.1016/j.jmr.2014.10.016, 2015. a, b

Walder, B. J., Conradi, M. S., Borchardt, J. J., Merrill, L. C., Sorte, E. G., Deichmann, E. J., Anderson, T. M., Alam, T. M., and Harrison, K. L.: NMR spectroscopy of coin cell batteries with metal casings, Sci. Adv., 7, eabg8298, https://doi.org/10.1126/sciadv.abg8298, 2021. a, b, c, d, e

Wapler, M. C., Leupold, J., Dragonu, I., von Elverfeld, D., Zaitsev, M., and Wallrabe, U.: Magnetic properties of materials for MR engineering, micro-MR and beyond, J. Magn. Reson., 242, 233–242, https://doi.org/10.1016/j.jmr.2014.02.005, 2014. a

Webster, R. D.: In situ electrochemical-NMR spectroscopy. Reduction of aromatic halides, Anal. Chem., 76, 1603–1610, https://doi.org/10.1021/ac0351724, 2004. a

Zhang, X. and Zwanziger, J. W.: Design and applications of an in situ electrochemical NMR cell, J. Magn. Reson., 208, 136–147, https://doi.org/10.1016/j.jmr.2010.10.013, 2011. a

Zhang, Z., Marble, A. E., MacGregor, R. P., Martin, J., Wang, H., and Balcom, B. J.: Zero-mode TEM parallel-plate resonator for high-resolution thin film magnetic resonance imaging, Can. J. Chem., 89, 745–753, https://doi.org/10.1139/v11-018, 2011. a, b, c

Download
Short summary
We developed a workflow using finite element methods to optimise electrochemical cell designs for in operando nuclear magnetic resonance by accurately matching magnetic field and radio frequency field simulations with experimental data. Guidelines for enhanced sensitivity and field homogeneity are given. A radio frequency amplification effect in coin cells is described by empirical formulae, which have the potential to improve spatial selectivity in future in operando applications.